DEVICE CONTROLS

STOP!

 Do not use the Multi-Mode Stimulator until you have read the 

entire instruction manual, especially the warnings and precautions on 

pages 5-6

.

STANDARD OPERATION INSTRUCTIONS

1. Begin by placing the electrodes on your skin at the treatment location. Follow 

your healthcare professional’s instructions as to the correct placement of the 

on the proper placement and attachment of the electrodes. Use care at this 
important step.

2. Make sure the stimulator is turned OFF. Connect the lead wires to the elec-

trodes and connect the opposite end to the therapy cable connector on the top 
of the stimulator.

3. Turn the stimulator ON. In a few moments, you will see a screen asking you 

to make a therapy selection. Using the “+” and “-” buttons scroll through the 
protocols until you locate the desired therapy protocol.

4. Press ENTER to select the therapy protocol. The therapy has begun but the 

amplitude is set at zero. Increase the amplitude to the level recommended by 
your healthcare professional. Use the “+” or “-” buttons to adjust the amplitude 
to your desired level.

DC Power Jack

Amplitude increase 
and decrease, also 
menu scroll buttons

OFF / PAUSE button

Therapy Cable 
Connector

LCD Display

ON / ENTER Button

DC Power Jack

Amplitude increase 
and decrease, also 
menu scroll buttons

OFF / PAUSE button

Summary of Contents for Multi-Mode Stimulator

Page 1: ...Carrier Frequency 4000 Hz Interference Frequency IF 4001 Hz to 4150 Hz Net Interferential Frequency Output Voltage 22 5 volts peak 0 45 volts peak to peak adjustable in 1 10 volt increments Output Cur...

Page 2: ...device KEEP THIS DEVICE OUT OF THE REACH OF CHILDREN Follow your physician s or therapist s instructions as to which therapy presets to use electrode placement and suggested amplitude levels TABLE OF...

Page 3: ...y that does change depending on the therapy protocol When these two stimulation frequencies cross and interfere with each other within the body the two frequencies subtract resulting in a third freque...

Page 4: ...an or therapist The amplitude or intensity of the treatment is patient controllable INTRODUCTION TO THE MULTI MODE STIMULATOR The Multi Mode Stimulator is a high quality advanced technology medical de...

Page 5: ...ns etc 7 Stimulation should not be applied over or in proximity to cancerous lesions 8 Stimulation should not be used whenever pain syndromes are undiagnosed until etiology is established PRECAUTIONS...

Page 6: ...Electronic monitoring equipment such as ECG monitors and alarms may not operate properly when interferential stimulation is in use in the management of pain patients 14 The Multi Mode Stimulator shou...

Page 7: ...tude Lockout To prevent an accidental increase or decrease of the treat ment amplitude intensity the Multi Mode Stimulator has an intra therapy amplitude lock feature Sixty seconds after the last incr...

Page 8: ...e quency for set dwell time then abrupt drop to base frequency repeat Relax and Contract Stimulation muscle stimulation Relax On for set dwell time at pre set base frequency and zero amplitude intensi...

Page 9: ...sec high 30 minutes 80 Hz 150 Hz 12 seconds Protocol Function Treatment Timer Base Freq High Freq Sweep Time Protocol 6 Full Sweep 6 seconds 30 minutes 1 Hz 150 Hz 6 seconds Protocol 7 Full Sweep 15 s...

Page 10: ...9 Dual MED Abrupt shift Ramped shift 15 minutes 15 minutes 1 Hz 80 Hz 10 Hz 150 Hz 6 seconds 6 seconds Protocol 10 Dual HIGH Abrupt shift Ramped shift 15 minutes 15 minutes 10 Hz 100 Hz 20 Hz 150 Hz 6...

Page 11: ...col 13 Muscle Simulation 20 minutes 50 Hz 50 Hz 6 sec ON 6 sec OFF Protocol 14 Muscle Simulation 20 minutes 50 Hz 50 Hz 5 sec ON 10 sec OFF Protocol 11 Protocol 12 Protocol 13 Protocol 14 3s 3s Contra...

Page 12: ...to the elec trodes and connect the opposite end to the therapy cable connector on the top of the stimulator 3 Turn the stimulator ON In a few moments you will see a screen asking you to make a therapy...

Page 13: ...your healthcare professional THERAPY STATUS SCREEN TYPICAL DISPLAY SCREENS SELECT PROTOCOL AMP LOCK PRESS ENTER TO UNLOCK THERAPY COMPLETE A090 P12 TR 07 Amplitude Setting A setting of 90 equals 9 vol...

Page 14: ...replacement lead wires Simulation Felt but the Device is Off Device is defective DO NOT USE please contact manufacturer for repair or replacement TYPICAL DISPLAY SCREENS Continued SYSTEM FAILURE CONTA...

Page 15: ...se Remove the batteries from the device during storage WARRANTY INFORMATION Elite Medical Supply of NY warrants the Multi Mode Stimulator against any defects in materials and workmanship for a period...

Page 16: ...Manufactured by Elite Medical Supply of NY LLC 1900 Ridge Road Suite 125 West Seneca NY 14224 866 712 0881 For questions regarding proper use or to request more supplies please contact Elite Medical S...

Reviews: