MEDICA - R200 US Version 01 08/14
Page 41
10.1
Sanitization selection (Process off)
1.
ENSURE the unit is in the process off mode.
2.
PRESENT sanitization PASSkey.
3.
PRESS TICK
.
4.
PRESS SCROLL
to type of sanitization required.
5.
PRESS TICK
.
10.2
RO sanitization (Process off)
1.
SELECT RO sanitization - ‘insert sanitant’ icon will
appear
OR
PRESS CROSS
to escape.
2.
REMOVE RO particle filter bowl (FH1) using filter
spanner.
3.
REMOVE filter (LC159) and discard if blocked or >6
months old.
4.
EMPTY remaining water out of filter bowl.
5.
ATTACH chemical tube (TUBE37548) to underside
of filter housing.
6.
POUR 150ml of Minncare into filter bowl and refit.
7.
TIGHTEN with filter spanner.
WARNING:
ALWAYS ENSURE SAFETY GLASSES AND
PROTECTIVE GLOVES ARE WORN WHEN
HANDLING THE SANITANT.
8.
PRESENT sanitization PASSkey to confirm sanitant
has been inserted.
Note: The unit will automatically enter the sanitization
mode. If the sanitization mode is stopped
through alarms or manual intervention then the
system will reiniate from the beginning of the
sanitization/ flush cycle.
9.
RO SANITIZATION takes approximately 1 hour.
Note: Once initated the unit will run for 45secs and then
enter a soak period of 45mins.
The final stage comprises a 15min. rinse.
RO sanitization screens
Chemical tube installation