ELGA
MEDICA-R 7/15 US
Operator Manual
MEDICA-R 7/15 US
Version 2
– 10/15
Page 19
Step 2 - UV lamp replacement timer
Setting this screen will cause the UV lamp timer to reset to the
preset value of 8760 hours (1 year).
1. PRESS tick
to jump to the next consumable or
to initiate reset.
2. PRESS tick
to reset timer or press
to abort
reset.
3. PRESS tick
to jump to the next consumable.
Step 3 - Pre-treatment cartridge replacement timer
Setting this screen will cause the pre-treatment cartridge timer
to reset to the preset value of 4380 hours (6 months).
1. PRESS tick
to jump to the next consumable or
to initiate reset.
2. PRESS tick
to reset timer or press
to abort
reset.
3. PRESS tick
to jump to the next consumable.
Step 4 - 0.2 µm filter replacement timer
Setting this screen will cause the 0.2µm filter replacement timer
to reset to the reset valve of 4380 hours (6 months).
1. PRESS tick
to jump to the next consumable or
to initiate reset.
2. PRESS tick
to reset timer or press
to abort
reset.
3. PRESS tick
to jump to the next consumable.
Pre-treatment Cartridge
Replacement Timer
UV Lamp Replacement Timer
0.2µm Filter Replacement
Timer