MEDICA-R 7/15 US
Operator Manual
ELGA
Page 14
MEDICA-R 7/15 US
Version 2
– 10/15
5.3 Connecting up the
MEDICA-R 7/15 US
Once the
MEDICA-R 7/15 US
unit has been positioned either on a
wall or on a bench, it should be connected as follows:
Mains water inlet tube
Drain
Pump feed
Outlet to analyzer
Outlet to reservoir/docking vessel
Step 1
– Fitting Tubes
1. PUSH in collet on connector.
2. PULL out transit plug.
3. CUT a clean square end on a 8mm (5/16") O/D
semi rigid drain tube.
4. PUSH tube into connector.
CAUTION! Do not restrict drain line.
CAUTION! If the water supply is at a pressure greater than
6 bar (90 psi) fit a pressure regulator (LA512).
Fitting Tubes
Delivery Pump and Drain
Connections
Outlet to Docking
Vessel/Reservoir
Outlet to
Analyzer
Pump
Feed
Feed Water Inlet
Connections
Blanking
plug
Drain
Feed
Water