5. COMMISSIONING SANITIZATION
5.1
Liquid Sanitization
5.1.1 RO Sanitization
Sanitization of the finished installation is essential to ensure
that the system is properly commissioned and capable of
achieving the bacteriological control required for a typical
application.
WARNING! ALWAYS WEAR RUBBER GLOVES, APRON
AND FACEMASK. DO NOT BREATHE FUMES
OR ALLOW TO COME IN CONTACT WITH SKIN
OR
EYES.
ALWAYS
FOLLOW
RECOMMENDATIONS
FOUND
IN
THE
MANUFACTURERS MATERIAL SAFETY DATA
SHEET
AND
ANY
APPLICABLE
OSHA
STANDARDS FOR THE CHEMICAL BEING
USED.
WARNING! LABEL THE MACHINE WITH APPROPRIATE
WARNING SIGNS SUCH A
S “DO NOT USE/
CONTAINS STERILANT” (NOT PROVIDED)
Minncare Cold Sterilant is a Peracetic and Hydrogen Peroxide
based solution.
Refer to the Minncare label for additional information and
follow
all applicable directions for use on the manufacturer’s
label in conjunction with the following instructions.
1.
ISOLATE application and ensure the process is OFF.
2.
PRESENT GREEN sanitization key to reader.
Sanitization selection screen
3.
PRESS
✓ to enter sanitization process.
Present the sanitization key