SYSTEM CONFIGURATION - PURELAB CHORUS 1 ANR/LSC/GSC
Chorus Port 1: Feedwater Inlet
Chorus Port 3: Drain
Chorus Port 4: Product Outlet
Chorus Port 5: Return
Chorus Port 6: Power
Chorus Port 7: Comms
Chorus Port 8: Power
Chorus Port 9: Comms
Chorus Reservoir Port 1: Inlet
Chorus Reservoir Port 4a: Outlet
Chorus Reservoir Port 4b: Outlet
Chorus Reservoir J1: Comms
Chorus Reservoir J2: Comms
Dispenser Port 1: Inlet
Dispenser Port 5: Return
Halo Dispenser Port 1: Inlet
Halo Dispenser Port 5: Return
Halo Dispenser Port 6: Power
Halo Dispenser Port 7: Comms
Halo Dispenser Port 8: Power
Halo Dispenser Port 9: Comms
Example 1: 2 X PURELAB Chorus 3 RO or 2 X PURELAB Chorus 2 RO/DI - 1 X PURELAB Chorus 1 - 3 X PURELAB Dispensers LA824
(
Max 4 Dispensers
) - LA759 60L Reservoir or LA760 100L Reservoir
®
®
®
®
3
4
6
9
7
7
7
7
9
5
5
5
5
Air Break Device
Air Break Device
Chorus 3 RO or Chorus 2 RO/DI fills an external reservoir with purified Type 3 or Type 2 water, which then feeds a Chorus 1
to produce Type 1 water.
New system configuration:
1. Commission Chorus 3 RO or Chorus 2 RO/DI systems first, directing reservoir port: 4a to drain during this procedure.
(Please refer to operator manual Chorus 3 RO MANU39996 or Chorus 2 RO/DI MANU40003 for commissioning instructions)
2. Once the reservoir is full, open reservoir port:4a manual valve to Chorus 1 port 1: Feedwater inlet and continue to commission Chorus 1
with the PURELAB Dispensers connected.
3. Chorus 1 is set in pre-commisioning mode and will automatically direct water to drain during this procedure.
(Please refer to operator manual Chorus 1 MANU39998 for commissioning instructions)
Existing system configuration:
If you are connecting a PURELAB Dispenser to an existing system configuration, please ensure all of your products
are up to date and using the latest software. To find the software version of your existing system, simply turn the power off by the wall and back on
again and the current software version number will be displayed in the bottom right-hand corner of the screen. If your software version number is
older than the one listed on the website at www.elgalabwater.com/customize, then please perform a software update,
see section 8.3 'Software Update' in this operator manual.
Potable Mains Feedwater Connection
Air Break
Device
4
J1
J2
4a
4b
4
3
3
8
6
6
Reservoir Overflow
COMMS
Key
Power
Water
Valve
15mm Valve
8mm Flow Bend
8mm Tube Clip
Regulator LA512
8mm Equal T
Dispenser Type 1
Dispenser Type 1
Dispenser Type 1
Chorus 3 RO or 2 RO/DI
Chorus 3 RO or 2 RO/DI
Reservoir 60L or 100L
Chorus 1
Type 3 or 2
Max Distance 5000mm
Max Distance 5000mm
®
Outlet to Application
Air Break Device
PURELAB
®
Dispenser