7.1 Replacing the Point-Of-Use Filter LC134 / LC145 / LC197
Step 1 - Removing the Dispense Tip
1. Unscrew the dispense tip from underneath the handset
(Fig.1)
.
2. Do not dispose of the dispense tip, this part is required for the sanitization procedure
(Fig.2)
.
Step 2 - Unpacking the LC134 / LC145 / LC197 Point-Of-Use Filter and Installing
1. Unpack the new POU filter from it's outer and inner packaging, remove the cap
(Fig.3)
.
2. Coupler is not required for PURELAB
®
Dispenser, screw in the Point-of-Use filter
(Fig.4)
.
3. Slightly open both bleed valves, before dispensing any water to release any trapped air
(Fig.5).
4. Press the dispense button, drawing water through and close both bleed valves
(Fig.6)
.
5. Installation is completed, normal operation can resume.
YOUR GUIDE TO THE CHORUS
7. MAINTENANCE
Dispense Tip
Only use 1 X
o-ring
Open
Close
Completed
(Fig.1)
(Fig.5)
(Fig.6)
(Fig.2)
Coupler
(not required)
Point-Of-Use LC134 / LC145 / LC197
Bleed Valve
M8mm Thread
Bleed Valve
Cap
(Fig.3)
PURELAB
®
Dispenser Operator Manual MANU41319 Version D 08/20 Page 22
(Fig.4)