4-2. Measurement Specifications

Accuracy of specifications apply for 1 year after purchase when
operated in a temperature of 18

O

C to 28

O

C (64

O

F to 82

O

F) and a

relative humidity of less than 75%.
Basic electrical specifications are given as +[(% of reading) +
(number of least significant digits)].

DC Voltage

200mV

100

m

V

+(0.25% of rdg+2dgt)

2V

1mV

+(0.8% of rdg+2dgt)

20V

10mV

+(0.8% of rdg+2dgt)

DC 1000V

200V

100mV

+(0.8% of rdg+2dgt)

or PEAK AC

1000V

1V

+(0.8% of rdg+2dgt)

NORMAL MODE REJECTION RATIO: greater than 46dB at 50Hz, 60Hz (1k unbalance)

Overload Protection: 1100VDC or peak AC for 50 seconds.

AC Voltage

200mV

100

m

V

+(1.0% of rdg+4dgt)

2V

1mV

+(1.0% of rdg+4dgt)

20V

10mV

+(1.0% of rdg+4dgt)

AC 750V max.

200V

100mV

+(1.0% of rdg+4dgt)

50Hz - 400Hz

750V

1V

+(1.5% of rdg+5dgt)

Resistance

200

W

0.1

W

+(0.8% of rdg+4dgt)

2k

W

1

W

+(0.8% of rdg+4dgt)

20k

W

10

W

+(0.8% of rdg+4dgt)

Protected by

200k

W

100

W

+(0.8% of rdg+4dgt)

PTC

2M

W

1k

W

+(1.5% of rdg+5dgt)

(500V rms)

20M

W

10k

W

+(2.0% of rdg+10dgt)

200M

W

100k

W

+(2.0% of rdg+10dgt)

DC Current

20

m

A

10nA

+(0.8% of rdg+4dgt)

200

m

A

100nA

+(0.8% of rdg+4dgt)

2mA

1

m

A

+(0.8% of rdg+4dgt)

Protected by

20mA

10

m

A

+(0.8% of rdg+4dgt)

0.5A/250V

200mA

100

m

A

+(1.0% of rdg+3dgt)

Fuse

10A

10mA

+(1.5% of rdg+3dgt)

10A/250V Fuse

Maximum Measuring Voltage Drop: Less than 300mV.

-6-

(4) Diode Test

1. Plug the red test lead into the “V/

W

/Hz” input jack and the black one into

the “COM” input jack.

2. Set the function rotary switch to                 position.
3. If the semiconductor junction being measured is connected to a circuit,

turn off the power to the circuit being tested and discharge all capacitors.

4. Connect the test leads to the device and read the forward value on the

digital display.

5. If the digital display reads overrange (OL), reverse the lead connections.

The placement of the test leads when the forward reading is displayed
indicates the orientation of the diode.
The red lead is positive and the black lead is negative. If overrange (OL)
is displayed with both lead connections, the junction is open.

(5) Transistor  h

FE

Measurement

1. Transistor must be out of circuit.
2. Set the function/rotary switch to the h

FE

position.

3. Plug the emitter, base and collector leads of the transistor into the correct

holes in either the NPN or the PNP transistor test socket, whichever is
appropriate for the transistor you are checking.

4. Read the h

FE

(beta or DC current gain) in the display.

(6) Capacitance Measurement

1. Set the function/rotary switch to the desired capacitance position.
2. Discharge capacitor before measurement.
3. Insert the capacitor leads into the capacitor test socket.
4. One lead must be inserted into one hole and the other lead must be

inserted into the other hole.

5. Read the capacitor value on the digital display.

(7) Continuity Test

1. Plug the red test lead into the “V/

W

/Hz” input jack and the black one into

the “COM” input jack.

2. Set the function/rotary switch to                 position.
3. Buzzer sounds if the resistance between two leads is less than approx.

30

W

.

(8) Frequency Measurement

1. Plug the red test lead into the “V.

W

.Hz” input jack and the black one into

the “COM” input jack.

2. Set the function/rotary switch to the “Hz” position.
3. Connect the test leads to the frequency source to be measured.
4. Read the frequency value on the digital display.

-11-

Range

Resolution

Accuracy

Maximum Input

Range

Resolution

Accuracy

Maximum Input

Range

Resolution

Accuracy

Input Protection

Range

Resolution

Accuracy

Input Protection

Summary of Contents for M-2785

Page 1: ...nter Avenue Wheeling IL 60090 847 541 3800 http www elenco com e mail elenco elenco com OPERATOR S INSTRUCTION MANUAL M 2785 WIDE MEASURING DIGITAL MULTIMETER with 15MHz Max Frequency Measurement Copy...

Page 2: ...set it on a flat surface Holding the Test Leads The tilt stand equips the two holes at the center to hold the test leads while you are not doing any measurements Simply set each red black lead in eac...

Page 3: ...features of the meter might be impaired Do not use the meter if the meter or the test leads look damaged or if you suspect that the meter is not operating properly Turn off the power to the circuit u...

Page 4: ...2 digit digital multimeter that is rugged reliable and convenient to use while providing all of the accuracy and features needed for any application It performs DC AC voltage DC AC current resistance...

Page 5: ...enance 12 3 Features 10 Functions 45 Measuring Ranges Surface Mounted PC Board Assembly 1 Basic Accuracy 10MW Input Impedance Large LCD with Function Range Annunciator Full Overload Protection Peak Ho...

Page 6: ...the forward value on the digital display 5 If the digital display reads overrange OL reverse the lead connections The placement of the test leads when the forward reading is displayed indicates the o...

Page 7: ...the COM input jack 2 Set the function rotary switch to the desired W position 3 If the resistance being measured is connected to a circuit turn off the power to the circuit being tested and discharge...

Page 8: ...perature variations 4 In order to prevent damage or injury to the unit never fail to keep the maximum tolerable voltage and current especially for the 10A current range 5 Carefully inspect the test le...

Reviews: