3 .1 .3 .1 Sin g le s t imu lu s a lg o rit h m
This algorithm is used to set the single stimulus test intensity w hen the
Threshold Related option is selected.
The threshold sensitivity is measured at four test locations, one in each
quadrant. The locations are displaced 9 degrees from the vertical and
horizontal meridians.
At each location the algorithm starts off 1dB brighter than the expected
threshold for the patient's age. It uses a repetitive bracketing procedure
w ith 1dB steps until six presentations have been made at all four
locations.
The average intensity of the last four presentations, at each of the four
test locations, is then taken as the threshold. To guard against the
inclusion of data from locations w here the threshold is abnormally
depressed, the algorithm excludes data from locations w here the
average of the last four presentations is below the 95% confidence
limits of the expected age setting. If all test locations are excluded, the
threshold is set at 4dB below the age setting.
presentations are made prior to the collection of
Heart data.
75
Summary of Contents for HENSON 8000
Page 1: ... 2017 Elektron Technology Henson Perimeter User Manual ...
Page 6: ......
Page 7: ...Part I ...
Page 14: ......
Page 15: ...Part II ...
Page 21: ...Part III ...
Page 25: ...25 Visual field tests 2017 Elektron Technology ...
Page 41: ...41 Visual field tests 2017 Elektron Technology ...
Page 64: ......
Page 65: ...Part IV ...
Page 85: ...Part V ...
Page 95: ...Part VI ...
Page 106: ......
Page 107: ...Part VII ...
Page 116: ......
Page 117: ...Part VIII ...
Page 121: ...Part IX ...
Page 123: ...Part X ...
Page 127: ...Part XI ...
Page 136: ......
Page 137: ...Part XII ...
Page 146: ......
Page 147: ...Part XIII ...
Page 149: ...Part XIV ...
Page 156: ......