Elektro-Automatik

Operating Guide

PS 9000 T

DC Laboratory Power Supply

Doc ID: PS9TEN
Revision: 01
Date: 04/2017

Summary of Contents for PS 9000 T Series

Page 1: ...Elektro Automatik Operating Guide PS 9000 T DC Laboratory Power Supply Doc ID PS9TEN Revision 01 Date 04 2017...

Page 2: ......

Page 3: ...2 2 3 8 Connecting the USB port rear side 32 2 3 9 Initial commission 33 2 3 10 Commission after a firmware update or a long period of non use 33 3 OPERATIONANDAPPLICATION 3 1 Personal safety 34 3 2 O...

Page 4: ...10 T 06200453 1 1 4 Symbols and warnings Warning and safety notices as well as general notices in this document are shown in a box with a symbol as follows Symbol for a life threatening danger Symbol...

Page 5: ...Construction not printed everywhere T Tower model Maximum current of the device in Ampere Maximum voltage of the device in Volt Series 9 Series 9000 Type identification PS Power Supply always programm...

Page 6: ...o equipment under power sparking may occur which can cause burns as well as damage to the equipment and to the source ESD regulations must be applied when plugging interface cards or modules into the...

Page 7: ...responsibilities for operation maintenance and cleaning of the equipment ensure that all employees who use the equipment have read and understood the manual Furthermore the users are to be regularly...

Page 8: ...device itself due to a defect Critical The device and or the load could be damaged Signal OCP OverCurrent Shutdown of the DC output due to excess of the preset limit Non critical protects the load fr...

Page 9: ...ge 0 352 W 0 352 W 0 352 W Output capacitance 4760 F 3400 F 720 F Temperature coefficient for set values K Voltage current 100 ppm Voltage regulation Adjustment range 0 40 V 0 80 V 0 200 V Accuracy 1...

Page 10: ...output U I Control signals DC on off remote control on off Status signals CV OVP OT Galvanic isolation to the device Max 1500 VDC Sample rate of inputs outputs 500 Hz Insulation Output DC to enclosure...

Page 11: ...4 W 0 704 W Output capacitance 4760 F 3400 F 720 F Temperature coefficient for set values K Voltage current 100 ppm Voltage regulation Adjustment range 0 40 V 0 80 V 0 200 V Accuracy 1 at 23 5 C 0 1 U...

Page 12: ...m side inlet rear exhaust Ambient temperature 0 50 C Storage temperature 20 70 C Humidity 80 not condensing Standards EN 61010 EN 61326 Overvoltage category 2 Protection class 1 Pollution degree 2 Ope...

Page 13: ...00 W 0 1100 W Output capacitance 6120 F 6120 F 1020 F Temperature coefficient for set values K Voltage current 100 ppm Voltage regulation Adjustment range 0 40 V 0 80 V 0 200 V Accuracy 1 at 23 5 C 0...

Page 14: ...mm side inlet rear exhaust Ambient temperature 0 50 C Storage temperature 20 70 C Humidity 80 not condensing Standards EN 61010 EN 61326 Overvoltage category 2 Protection class 1 Pollution degree 2 Op...

Page 15: ...66 A Overpower protection range 0 1100 W 0 1650 W 0 1650 W Output capacitance 130 F 6120 F 6120 F Temperature coefficient for set values K Voltage current 100 ppm Voltage regulation Adjustment range 0...

Page 16: ...0 mm side inlet rear exhaust Ambient temperature 0 50 C Storage temperature 20 70 C Humidity 80 not condensing Standards EN 61010 EN 61326 Overvoltage category 2 Protection class 1 Pollution degree 2...

Page 17: ...ion range 0 220 V 0 550 V Overcurrent protection range 0 27 5 A 0 11 A Overpower protection range 0 1650 W 0 1650 W Output capacitance 1020 F 130 F Temperature coefficient for set values K Voltage cur...

Page 18: ...de inlet rear exhaust Ambient temperature 0 50 C Storage temperature 20 70 C Humidity 80 not condensing Standards EN 61010 EN 61326 Overvoltage category 2 Protection class 1 Pollution degree 2 Operati...

Page 19: ...33 41747 Viersen Germany Fon 49 2162 3785 0 Fax 49 2162 16230 www elektroautomatik de ea1974 elektroautomatik de PS 9000 T Series A Control panel with display B Knobs C USB port D DC output E Remote...

Page 20: ...automatik de ea1974 elektroautomatik de PS 9000 T Series Figure 2 Rear view 320 W 640 W version shown Figure 3 Rear view 1000 W 1500 W version shown F Control interfaces digital analog J Power switch...

Page 21: ...EA Elektro Automatik GmbH Helmholtzstr 31 33 41747 Viersen Germany Fon 49 2162 3785 0 Fax 49 2162 16230 www elektroautomatik de ea1974 elektroautomatik de PS 9000 T Series Figure 4 Top view Figure 5 S...

Page 22: ...oltage set value or setting the parameter values in the menu Left knob push selection of the decimal position to be changed cursor for the assigned value Right knob turn adjusting the current or power...

Page 23: ...standard with a USB port on the back side which can be optionally extended to a 3 way interface with USB Ethernet and a galvanically isolated analog port all retrofittable by the user The configuratio...

Page 24: ...Schuko plug or UK plug depending on shipping destination 1 x USB cable 1 8 m 1 x USB stick with software and documentation 1 9 4 Optional accessories The below listed optional accessories can be purc...

Page 25: ...nd power are displayed When the DC output is switched on the actual regulation mode CV CC or CP is displayed in the corresponding area like depicted in the figure above with CC in the red area for cur...

Page 26: ...on on page 37 1 9 5 3 Pushbutton function of the knobs The rotary knobs also have a pushbutton function which is used in all menu options for value adjustment to move the cursor by rotation as shown 1...

Page 27: ...a recorded during normal operation in all modes The file layout is identical to the those generated from the Logging feature in EA Power Control The nr field in the file name is automatically counted...

Page 28: ...ssion speed is set to Auto negotiation and means it can use 10MBit s or 100MBit s 1GB s is not supported Duplex mode is always full duplex If remote control is going to be activated the Ethernet inter...

Page 29: ...Installation 2 3 1 Safety procedures before installation and use Before connecting to the mains ensure that the supply voltage is as shown on the product label Overvoltage on the AC supply can cause...

Page 30: ...de ea1974 elektroautomatik de PS 9000 T Series 2 3 3 1 Placement on horizontal standing surfaces The device is designed as a desktop unit and must only be operated on horizontal surfaces which are cap...

Page 31: ...e cables are long then the cross section must be increased to avoid voltage loss and overheating 2 3 4 1 Possible connections on the DC output The DC output on the front is of type clamp plug and can...

Page 32: ...The optionally available 15 pole connector Type Sub D D Sub on the rear side is an analog interface To con nect this to a controlling hardware PC electronic circuit a standard plug is necessary not i...

Page 33: ...ables to be used are of a satisfactory cross section Check if the factory settings of set values safety and monitoring functions and communication are suitable for your intended application of the dev...

Page 34: ...tion Constant voltage Voltage regulation is also called constant voltage operation CV The DC output voltage of a power supply is held constant on the adjusted value unless the output current or the ou...

Page 35: ...t power CP keeps the DC output power of a power supply constant if the current flowing to the load in relation to the output voltage and the resistance of load reaches the adjusted value according to...

Page 36: ...plying power temporarily After cooling down the device will automatically continue to supply power while the condition of the DC output remains and the alarm doesn t require to be acknowledged 3 3 3 O...

Page 37: ...r off All set values are always saved and restored 3 4 2 Switching off the device On switch off the last output condition and the most recent set values are saved Furthermore a PF alarm power failure...

Page 38: ...s Overview Page 1 Page 2 Page 7 Enter Page 6 Output Settings Protection Settings OVP OCP OPP Limit Settings General Settings Reset device Allow remote control Yes No DC output after power ON OFF Resto...

Page 39: ...like with U I for Sine where U A changes to I A etc MENU Communication About HW SW IP Settings 1 IP Settings 2 Logging Enable USB logging Com Timeout Com Protocols Timeout USB RS232 ms 5 65535 SCPI M...

Page 40: ...974 elektroautomatik de PS 9000 T Series Default profile User Profile 1 User Profile 2 User Profile 3 User Profile 4 User Profile 5 MENU Profiles Backlight English Deutsch Language 60 sec on Always on...

Page 41: ...as local in the status area on the main display See also section 1 9 5 1 DC output after power ON 1 Determines the condition of the DC output after power up OFF DC output is always off after switching...

Page 42: ...delivery the device only features an USB port which doesn t require configuration It can be extended by an Ethernet LAN port by installing the optional 3 way interface board IF KE4 After installation...

Page 43: ...B stick feature Once enabled you can define the log interval multiple steps 500 ms 5 s and choose between Start stop with DC on off or Manual start stop With a properly formatted USB stick also see 1...

Page 44: ...tary knob is made for 60 s As soon as input is made the backlight returns automatically Furthermore the brightness can be selected in 10 steps HMI Lock See 3 7 Control panel HMI lock on page 55 Status...

Page 45: ...the touch area for direct input is tapped bottom center The adjustment limits are coupled to the set values It means that the upper limit may not be set lower than the corresponding set value Example...

Page 46: ...will switch the knob to power adjustment and highlight the power value 3 After successful selection the desired value can be set within the defined limits Selecting a digit is done by pushing the rot...

Page 47: ...standard status page mode with knobs or by direct input Regulation mode CP is displayed alternatively to CC at the same position 3 4 7 Switching the DC output on or off The DC output of the device ca...

Page 48: ...ting Manual start stop it is different Logging is then started and stopped only in the MENU in the logging configuration page Soon after logging has been started the symbol indicates the ongoing loggi...

Page 49: ...otely Activating condition Local causes the following If remote control via the digital interface is active Remote then it is immediately terminated and in order to continue remote control once Local...

Page 50: ...set values VSEL CSEL PSEL correspond to 0 100 nominal values 0 100 actual values correspond to 0 5 V at the actual value outputs CMON VMON 0 10 V Reference voltage 10 V 0 10 V set values VSEL CSEL PSE...

Page 51: ...ax 5 mA Short circuit proof against AGND 4 DGND POT Ground for all digi tal signals For control and status signals 5 REMOTE DI Switching internal remote control Remote LOW ULow 1 V Internal HIGH UHigh...

Page 52: ...REM SB A digital output e g from a PLC may be unable to cleanly pull down the pin as it may not be of low enough resistance Check the specification of the controlling application Also see pin diagram...

Page 53: ...ocked It can be switched on or off by pushbutton or digital command LOW inverted HIGH inverted DC output will be switched off and locked Later it can be switched on again by toggling the pin During lo...

Page 54: ...eshold The DC output will be switched off 0 V 1 1 UNom Display analog IF digital IF OCP OverCurrent Protection Triggers an alarm if the DC output current reaches the defined threshold The DC output wi...

Page 55: ...e HMI by tapping on Tap to unlock within 5 seconds otherwise the pop up will disappear and the HMI remains locked In case the additional PIN code lock has been activated in the menu HMI Lock another r...

Page 56: ...ser profile to work with 1 While the DC output is switched off tap the touch area on the main screen 2 In the menu page tap on Page 2 and then on Profiles 3 In the selection screen see image to the ri...

Page 57: ...urrent and hence higher power This can be achieved by connecting all units to the DC load in parallel so the single currents can add There is no support for a balancing between the individual units li...

Page 58: ...the supplier has been contacted and it is clarified how and where the equipment should be sent the device is in fully assembled state and in suitable transport packaging ideally the original packaging...

Page 59: ...t conditions are required A measurement device multimeter for voltage with a max error of half the PS s voltage error That measure ment device can also be used to measure the shunt voltage when readju...

Page 60: ...correct and submit with ENTER 5 Repeat step 4 for the next three steps total of four steps In case you are generally using the remote sensing feature it is recommended to also readjust this parameter...

Page 61: ...epairs if not otherwise arranged between supplier and customer will be carried out by the manufacturer For this the device must generally be returned to the manufacturer No RMA number is needed It is...

Page 62: ......

Page 63: ......

Page 64: ...EA Elektro Automatik GmbH Co KG Development Production Sales Helmholtzstra e 31 37 41747 Viersen Germany Fon 49 2162 37 850 Fax 49 02162 16 230 ea1974 elektroautomatik de www elektroautomatik de...