Page 69
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
EL 9000 B 15U/24U Series
►
How to save a sequence point table to an USB stick:
1.
Do not plug the USB stick yet or remove it.
2.
Access the function selection menu of the function generator via MENU -> Function Generator -> Arbitrary
3.
Tap on
, then
. The device will request you to plug the USB stick now.
4.
After plugging it, the device will try to access the stick and find the folder HMI_FILES and read the content.
If there are already WAVE_U or WAVE_I files present, they will be listed and you can either select one for
overwriting with
, otherwise select
for a new file.
5.
Finally save the sequence table with
.
3.10.11
Ramp function
The following parameters can be configured for a ramp function.
Value
Range
Description
Ustart / Istart
0...Nominal value of U, I
Start value (U,I)
Uend / Iend
0...Nominal value of U, I
End value (U, I)
t1
0,1 ms...36000 s
Time before ramp-up or ramp-down of the signal
t2
0.1 ms...36000 s
Ramp-up or ramp-down time
Schematic diagram:
Application and result:
U
(I)
sta
rt
t
U,I
t1
t2
U
(I)
End
This function generates a rising or falling ramp between start and
end values over the time
t2
. Time
t1
creates a delay before the
ramp starts.
The function runs once and stops at the end value. To have a
repeating ramp, function Trapezoid would have to be used instead
(see
).
Important to consider are the static values of U and I which define
the start levels at the beginning of the ramp. It’s recommended
that these values are set equal to those in
Ustart
/
Istart
, unless
the source at the DC input should not be loaded before the start
of the ramp. In that case the static values should be set to zero.
10h after reaching the ramp end, the function
will stop automatically (i.e. I = 0 A resp. U =
0 V), unless it has been stopped manually
before.
3.10.12
UI and IU table functions (XY table)
The UI and IU functions offer the user the possibility to set a DC input current dependent on the DC input voltage,
or a DC input voltage dependent on the DC input current. The function is table driven with exactly 4096 values,
which are distributed over the whole measured range of actual input voltage or input current in the range of 0...125%
Unom or Inom, of which only 0...102% are effective. The table can either be uploaded from an USB stick through
the front side USB port of the device or via remote control (ModBus protocol or SCPI). The functions are:
UI function:
U = f(I) -> the device will work in CV mode, regulating the voltage of a source in CC mode
IU function:
I = f(U) -> the device shall work in CC mode with a source that is in CV mode
In the
UI function
, the equipment measuring circuit determines the level from 0 to maximum of the input current.
For each of the 3277 effective values for input current a voltage value is maintained by the user in the UI table which
can be any value between 0 and nominal value. The values uploaded from an USB stick will always be interpreted
as voltage values even if the user calculated them as current values and incorrectly loaded them as an UI table.
In the
IU function
the assignment of the values is the other way round, the behavior, however, the same.
Thus the behavior of the load or the DC current and power consumption can be controlled with dependence on
input voltage and step changes can be created.