5 • Calibration menu
●
13
Confirm your choice by pressing the [OK] button.
Press the [EXIT] button to leave all settings unchanged.
Choose your language (English, German, French, Dutch) in the corre-
sponding menu.
• to the calibration menus by a long press (the display responds to
the long press by flashing briefly).
16f. Use the rotary knob to navigate through the menus.
Confirm your choice by pressing the [OK] button.
Press the [EXIT] button to leave all settings unchanged.
To get accurate measurements, the measurement circuitry must be
calibrated. It is recommended to perform the calibration steps in the
order of the menus.
Note:
The menu option Change display: INVERSE/NORMAL (fig. 16e)
which allows the display to be inverted (light/dark background) only
appears once version 3 of the file LCR6 texts v3.txt has been loaded
into the device. This update (see also §9.2.5) is described in the man-
ual LCR Meter AU2019 - EN Operating Instructions (§4.2.4 et §4.2.5).
5 • Calibration menu
All the steps for calibrating the instrument are accessible through this
menu. An external multimeter (
not included in the kit
) is required. Preci-
sion resistors (0.01% tolerance) fitted on the PCB for your conveni-
ence get connected to the measurement input via one of the jumpers
on J10 to J13.
At each step, a message on the screen indicates the state required for
various jumpers (i.e. closed or open).