Type BF applied part:
Electrodes
IP22
Continuous operation
Electro-Medic device
internal power supply
Applied part
Protection of electrical
equipment against water
or fine particle infiltration
that may have a harmful
effect on the equipment.
Instructions
Note: Not intended to be sterilized.
Do not use the device in an oxygen-rich
environment.
Protection against
electrical overload
SAFETY CLASSIFICATION
OF ELECTRO-MEDIC EQUIPMENT
TIME TO RECHARGE
•
A Li-ion battery requires approximately 3.5 hours to recharge.
LI-ION BATTERY
•
Voltage limited to 4.2 V
•
Rechargeable 3.7 V/600 mAh Li-ion battery
ADAPTOR
•
100-240 V 50-60 Hz, 1.2 A
CAUTION
•
This equipment must not be connected with an adaptor other than that supplied
with the Electro-Medic equipment.
SAFETY MEASURES
•
Do not short-circuit the device
•
Do not expose the device to high temperatures
•
Use only the specifically recommended charger
The adaptor is 2MOPP equipment under IEC 60601-1-1.
Equipment approval is valid if used in combination with
the adaptor supplied with this equipment.
Do not position the device where access to
the main source of the device would prevent
the device from being shut down quickly
if necessary.