ELECTROMATIC WTT-110 User Manual Download Page 10

– 10 –

10.0 A

DJUSTING

A

DVANCED

S

ETTINGS

There are several adjustable settings that can be changed from the factory defaults to
customize the WTT-110 to meet the exact requirements of the user. These settings are
saved in the microprocessor and are recalled each time the system is powered up.

10.1 Temperature Compensation (Default: SET=on)

In general, it is recommended to have the this function turned on as long as it
does not effect your measuring results. Only when measuring very minute 
forces over a longer period of time at a slow rate and under stable environmental
conditions might it be useful to switch this off. 

Switch off the instrument.

Press and hold PEAK- and MODE-buttons.

Press ON/OFF-button.

Wait until TRK oFF is displayed.

Release PEAK- and MODE-buttons.

10.2 Selecting the baud rate of the RS232C interface (Default: 2400)

On the WT-110RS model ONLY, the speed of the data transfer for the RS232C
interface can be adjusted as follows: 

Switch off the instrument.

Press and hold RESET-button.

Press ON/OFF-button.

Wait until f01 is displayed, then release RESET-button.

Select function f04 by repeating to press the PEAK-button until f04 is shown 
on the display.

Select appropriate baud rate (2400/4800/9600/19200) by repeating to press 
the MODE-button.

Press RESET-button to save your settings.

10.3 Switching off Auto Power-Off function (Default: 10 min)

While operating under battery power the instrument will be automatically
switched off after approx. 10 minutes on inactivity. This Auto Power-Off func-
tion can be deactivated, which may become necessary when monitoring the 
fluctuation of applied forces for a longer period of time. This function is 
disabled when the instrument is powered by the AC-adaptor.

Switch off the instrument.

Press and hold RESET-button.

Press ON/OFF-button.

■ 

Wait until f01 is displayed, then release RESET-button.

Select function f03 by repeating to press the PEAK-button until f03 is shown 
on the display.

Select setting (10/OFF) by repeating to press the MODE-button.

Press RESET-button to save your settings.

TRK

oFF

– 11 –

10.4 Change the refresh rate of the display (Default: 3/sec)

The force applied to the load cell of the instrument is internally processed at a
rate of 1000 Hz, which allows the precise capturing of the peak values. The read-
ings on the display, however, are refreshed 3 times/sec only for the convenience
of the human eyes. You may increase or decrease the factory settings for them to
match your personal demands.

Switch off the instrument.

Press and hold RESET-button.

Press ON/OFF-button.

■ 

Wait until f01 is displayed, then release RESET-button.

Select function f02 by repeating to press the PEAK-button until f02 is shown 
on the display.

Select refresh-time (1/2/3/5/10/20 times/sec) by repeating to press the 
MODE-button.

Press RESET-button to save your settings.

Summary of Contents for WTT-110

Page 1: ...pair disassembly tampering abuse or misapplications has taken place as determined by ELECTROMATIC All returns for warranty or non warranty repairs and or replacement must be authorized by ELECTROMATIC in advance with all repacking and shipping expenses to the address below to be borne by the purchaser THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES EXPRESSED OR IMPLIED INCLUDING BUT NOT ...

Page 2: ...lder pins or similar wire terminal components in the field of quality control or design validation Safety Precautions Wear appropriate eye protection at all times The load cell can be damaged if the internal measuring system is overloaded Do not exceed the maximum measuring limit of 50 Kg 110 lbs or 500N Transport and store the instrument with great care This reduces the risk of damage to the load...

Page 3: ...tion 0 01Kg 0 1 lbs 0 1 N Terminal adapter slot width mm 0 5 0 8 1 0 1 4 1 5 2 0 2 5 3 0 3 5 4 0 5 0 6 0 Accuracy 0 5 F S or better Operating Mode Continuous Displays actual value in Kg lbs or N Peak Hold Displays peak value in Kg lbs or N Wire Diameter SAE AS7928 II AWG 8 30 IEC 60352 2 Cross section 0 05 10mm2 Maximum 0 236 6mm Overload 200 Full Scale LCD indicator at 120 Display LCD 4 1 2 digit...

Page 4: ... be operated using the AC Adapter Charger The batteries are being recharged while plugged into the AC power source Total battery recharge time will be significantly longer then if recharged while turned off 17 13 0 TROUBLESHOOTING GUIDE If any error codes are shown on the display first try turning the power off and then back on again If the code is still on the display refer to the chart below AWG...

Page 5: ... and ZERO until the smaller characters at the top of the display show CAL 4 Release the UNITS PEAK and ZERO keys 5 The force gauge is now in calibration mode 6 Press the UNITS switch The display will show ZER after blinking SCN for 10 seconds 7 The force gauge is now ready for zero point calibration 8 Press ZERO key to confirm the zero calibration Wait approximately 15 20 seconds The display will ...

Page 6: ...nit is handled with care and operated as detailed in this manual it should remain accurate for an extended time period If however it is subjected to forces that exceed its maximum range or if it is not properly cared for it might need to be recalibrated It is recommended that the calibration is verified at least on an annual basis and more frequently if feasible Normally instruments of this type g...

Page 7: ... for clearing the currently stored Peak value from display memory whenever desired 4 Select the smallest suitable slot in the Wire Terminal Fixture based on the diameter of the wire and the physical dimensions of the connector Rotate the fixture so the selected slot is in the 3 o clock position so it is closest to the Wire Clamp Fixture 5 Place the cable connection into the fixture such that the m...

Page 8: ... 50 sec BB3cr Data output request 100 sec BDcr Units confirmation request BEcr Peak data output request BFcr Minus peak data output request cr carriage return Chart continues on next page 11 0 DATA TRANSFER MODEL WTT 110RS ONLY Model WTT 110RS testers can transfer measuring data by means of an RS 232C inter face A 9 pin D Sub connector is provided for this purpose which can be found at the left si...

Page 9: ...mperature Compensation The WTT 110 utilizes a built in temperature compensation system to provide optimum accuracy and to prevent signal drift If the user would like to disable this feature please refer to the Advanced Settings section on page 10 12 OFF ZERO Press and hold ZERO button Press and release POWER button F01 0001 F01 0001 F01 0001 F02 3 F03 10 F04 2400 Plus or minus sign PEAK UNIT Data ...

Page 10: ...3 Switching off Auto Power Off function Default 10 min While operating under battery power the instrument will be automatically switched off after approx 10 minutes on inactivity This Auto Power Off func tion can be deactivated which may become necessary when monitoring the fluctuation of applied forces for a longer period of time This function is disabled when the instrument is powered by the AC ...

Page 11: ...I P M E N T C O I N C 600 Oakland Ave Cedarhurst NY 11516 U S A TEL 516 295 4300 FAX 516 295 4399 CHECK LINE INSTRUMENTS WTT 110 WTT 110RS WIRE TERMINAL PULL TESTER Operating Instructions CHECK LINE BY ELECTROMATIC OI311 ...

Reviews: