Electromatic Equipment Co., Inc. 

-4- 

2.0 

CALIBRATION CURVES 

 
 

Name of Calibration 

curve 

Material under test 

Fill quantity 

300g corn stand. 

corn standard 

300g 

300g corn hand 

corn hand reaped 

300g 

300g rye 

rye 

300g 

300g triticale 

triticales 

300g 

300g whea/spe.pe 

wheat / spelt peeled 

300g 

300g barley 

barley 

300g 

190g oats 

oats 

190g 

300g rape 

rape 

300g 

230g pumpkin 

pumpkin seed 

230g 

310g peas 

peas 

310g 

300g soybean 

soybean 

300g 

277g beetle beans 

beetle bean 

277g 

295g horse bean 

horse bean 

295g 

180g sunflower 

sunflower seed 

180g 

reference 

-------- 

-------- 

 

  Reference: To test the humimeter FS2. Do not use for measuring the moisture! 

 
 

 

Electromatic Equipment Co., Inc. 

-9- 

6.0 

CHANGING BATTERIES 

 
 
Your new device is provided with batteries. 
 
Change the batteries: 
 
1.   Press  with  your  finger  onto  the  arrow  of  the  battery 

cap and pull it back. 

 

2.   Remove the empty batteries. 

 

3.   Put  four  new  batteries  in  the  device.  Make  sure  the 

position of the battery poles is correct. 

 

4.   Press down the batteries and close the cap.  
 
 
 
 

7.0 

DETERMINATION OF THE MATERIAL MOISTURE 

 
 

The principle is a comparison measurement with the drying out method according to 
DIN 10350: 1967 09. Take the measured sample and weigh it. Dry it out in an oven 
and  weigh  it  again.  Following  formula  can  be  taken  to  determine  the  absolute 
moisture: 

 

Note:

 The dried material may not be used any more for measuring purposes! 

 
 
 

 

 
M

: Mass with average moisture content 

M

t

   : Mass of the dried sample 

%F  : Calculated absolute moisture 
 
 

Exemption from liability 

 
For miss-readings and wrong measurements and of this resulting damage we refuse 
any  liability.  This  is  a  device  for  quick  determination  of  moisture.  The  moisture 
depends on multiple conditions and multiple materials. Therefore we recommend a 
plausibility  check  of  the  measuring  results.  Each  device  includes  a  serial  number 
and  the  guarantee  stamp.  If  those  are  broken,  no  claims  for  guarantee  can  be 
made. In case of a faulty device, please contact Electromatic. 
 

 

%F =

               

x 100

  

M

n

 - M

M

Summary of Contents for FS2

Page 1: ... 1 0 Introduction 2 2 0 Calibration curves 4 3 0 Design of the device 5 3 1 Menu 3 2 Keypad symbols 4 0 Most common reasons for miss readings 8 5 0 Device maintenance instructions 8 6 0 Changing batteries 9 7 0 Determination of the material moisture 9 8 0 Technical data 10 10 0 Warranty 11 ...

Page 2: ...ust be authorized by Electromatic in advance with all repacking and shipping expenses to the address below to be borne by the purchaser THE FOREGOING WARRANTY S IN LIEU OF ALL OTHER WARRANTIES EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO THE WARRANTY OF MERCHANTABILITY AND FITNESS FOR ANY PARTICULAR PURPOSE OR APPLICATION ELECTROMATIC SHALL NOT BE RESPONSIBLE NOR LIABLE FOR ANY CONSEQUENTIAL ...

Page 3: ...app 1 3 kg with batteries Degree of protection IP 40 Scope of supply Plastic case Digital scale 1000g Measuring cup 0 5 litre 4 x 1 5Volt AA Alkaline Batteries Electromatic Equipment Co Inc 3 8 Fill up the measuring device with the sample material The filling needs to be done slowly and constantly to ensure reproducible results 9 The display shows the measuring result 10 To save the results in the...

Page 4: ... device Make sure the position of the battery poles is correct 4 Press down the batteries and close the cap 7 0 DETERMINATION OF THE MATERIAL MOISTURE The principle is a comparison measurement with the drying out method according to DIN 10350 1967 09 Take the measured sample and weigh it Dry it out in an oven and weigh it again Following formula can be taken to determine the absolute moisture Note...

Page 5: ... calibration curve Wrong fill quantity Fill exactly the right weight 1 0 g in the measuring chamber Wet or mouldy material Frozen measuring material 5 0 DEVICE MAINTENANCE INSTRUCTIONS To provide a long lives of your device please do not expose it to strong mechanical loads or heat e g dropping it or direct sunlight exposure Clean your device using a dry cloth The measuring chamber needs to be cle...

Page 6: ...t Log All Logs Clear Send Logs Manual Logs Clear Logs Options Date Time Language Unlock C F Password Reset Materialcalib Status Electromatic Equipment Co Inc 7 3 2 Keypad symbols Measuring window Rolling Menu Power ON OFF Switch upper Switch lower Save Hold Watch the measuring rows Suppliers data can be added Menu Enter Switch upper Switch lower Exit Enter numbers Enter letters Next or right Left ...

Page 7: ...FS2 GRAIN MOISTURE METER OPERATING INSTRUCTIONS Distributed by ABQ Industrial LP USA Tel 1 281 516 9292 888 275 5772 eFax 1 866 234 0451 Web https www abqindustrial net E mail info abqindustrial net ...

Reviews: