background image

PLUS

    PLUS-P

PLUS-485        PLUS 4-20mA
PLUS-232

English

L1

P2

N

L3

L2

P1

S2

S1

P2

S2

S1

P1

P2

S2

S1

P1

L1

N

L3

L2

50 mA T

200 ÷ 240 VAC

5   6

1  2  3  4

8  9 10

1 2 3 4

20 19 18 17

11 12

12 11

The PLUS range  of products are manufactured by Akse S.r.l. in compliance  with
EMC 89/336/EEC regulations and in accordance with the following standards:
EMISSIONS = EN 50081-1  1992 EN5022 CLASS  B, CISPR 22;  IMMUNITY=
EN 50082-1 1992 (light industry)   SAFETY=EN61010.

1

SAFETY

This instrument was manufactured  and tested in compliance with class  2  IEC 1010 and VDE
411 standards, in accordance with group B   VDE 0110 standards for operating voltages up to
250 VACrms phase to neutral. To maintain this condition and  to ensure safe operation, the user
must comply with the  indications and markings  contained in the  following instructions:
- when the instrument is received, before beginning installation, check that it is still intact and no

damage was incurred during the transport;

- before mounting, ensure that the operating voltage and mains voltage set are the same and then

proceed with the installation;

- the power supply does not require  an earth connection;
- a 50 mA T fuse should be installed in the power supply  circuit to the instrument;
- maintenance and/or repairs must be carried on only by qualified, authorized personnel;
- if there is ever the suspicion that safe use is no longer possible, the instrument must be discon-

nected and precautions must be taken against accidental use;

- operation is  no longer safe when:

1) there are clearly visible damages;
2) the instrument no longer functions;
3) after a prolonged storage in unfavorable conditions;
4) after serious damages incurred during transport.

1.1

OPERATOR SAFETY

Read these pages carefully before installing and  utilizing the instrument.

The instrument descibed in this user manual is intended for use by properly trained staff only.
Maintenance and/or repairs  must be carried out by authorized personnel only. For proper, safe use
of the  instrument and for maintenance and/or repair, it is essential that the persons instructed  to
carry out these  procedures follow normal safety precautions.

1.2

SYMBOLS

THIS  SYMBOL MEANS: "READ THE INSTRUCTIONS"

2

POWER SUPPLY

ATTENTION:

The instrument must have a power supply with voltage ranging from 200÷240 VAC 50/60 Hz
using max. cable  gauge 2.5  mm

2  

and attached to the power supply terminals (see Fig. below).

The instrument  power supply does not need an earth connection.  The instrument  requires the
installation of an external 50 m AT type fuse in the power supply circuit.

2.1

CONNECTION OF THE VOLTAGE INPUT

Use cables with max. gauge 2,5 mm

2

 and attach them to the screw terminals for voltage output.

The following drawings show how to connect the screw terminals to the phases. Refer to the
detailed wiring diagram to make the correct connection of the voltage input depending on the
selected  instrument  and on the plant it is installed into.

Rev. 06/01

Max 250 VAC

Phase-Neutral

2.2

CONNECTION OF THE CURRENT INPUT

The  instruments  are  equipped with terminals  for connection of the current signal inputs.  Connect
cables as indicated in the following diagrams.
The P1 and  P2 indications identify the current correct direction.

N.B.: 

 In case the CT connection is accidentally reversed, the instrument provides  an

automatic compensation and, hence, inputs will  be correct.

Above feature applies to  reversed voltage connection too.

It is, however, absolutely essential to follow the sequence of the phases   between voltage and

current.

N.B.  The secondary CT must  always short  circuit  when it is not  connected to the instrument

to avoid damages and risks to  the operator.

3

INSTRUMENT PROGRAMMING

The writing "WAIT"  appears  when the instrument is powered. After a few seconds the  measure-
ment will be displayed (the LED on the key is lighted).
The programming key

 is located at the bottom left corner of the front panel. To enter into

programming mode, press  simultaneously the key 

 and  the key 

. On the display the

CT selection will appear:

Pressing   the   key

 the available CT ratio is selected (see diagram in the following page)

Once the selection has been operated, press the programming key 

to confirm/exit set up.

Pressing simultaneously  the key 

 and  the key 

 the kWh values are zeroed.Pressing

simultaneously  the key 

and  the key 

 the kVArh values are zeroed.

The CT ratio prompted by default is  5/5.
Once the  instrument has  been programmed it is  possible  to disable the key 

 by removing the

jumper placed  near the current current inputs.
Once the jumper is removed it will not be possible to  reset the counters and  to change the CT ratio.
To enable again the programming functions, it is necessary to restore  the jumper.

Remove the jumper for disabling set up and reset

3.1

ADDRESS OF THE INSTRUMENT

 (only for models 485/232)

After entering the programming page through 

  and 

, press the key 

 once more and

the address of the instrument will be displayed. The default address set by the manufacturer upon
delivery is:

.

It can be changed through: key 

to select the  digit

Key 

 to increase the digit

Key 

to decrease the digit

The flashing digit is active.
To exit the programming, press key 

 a fourth time.

The maximum selectable number is 247.

AKSE S.r.l. via Aldo Moro, 39 - 42100 Reggio Emilia (RE) - Italy - Tel. +39.0522.924244  Fax. +39.0522.924245

Reviews: