Portable Conductivity/TDS Meter
513-160
14
9907X0094 Issue 2
3.5 Setup menu navigation
For general navigation in the setup menu read the following information:
Press and hold
Setup
to enter the setup menu.
Press
Exit
to exit the setup menu.
•
Use ˄ and ˅ to increase or decrease values.
Press
Read
to confirm a change.
The following parameters can be changed in the order as shown.
Parameter
Description
Range
MTC
Manual temperature setting
0.0…100.0
°
C /
32.0…212
°
F
Calibration standard setting
1, 2, 3
Ref.T.
Reference temperature
25
°
C (68
°
F),
20
°
C (77
°
F)
α
α
-coefficient
0.0…10.00 %/
°
C
TDS
TDS factor
0.4…1.00
°
C,
°
F
Temperature unit
°
C,
°
F
3.6 Measurement modes
With the conductivity meter it is possible to measure the following
parameters of a sample:
•
Conductivity (µS/cm and mS/cm)
•
TDS (mg/L and g/L)
To change the unit, press
Mode
on the measurement screen until the
desired appears.