Calibration wizard. Step 5.
New calibration constant is now calculated
Click button
Save
to save the new calculated
calibration constants.
Click button
Cancel
to cancel the guide without
updating the calibration constant.
•
Calibration wizard. Step 6.
Enter password and calibration signature.
See
7.3.6 Enter Password.
•
Calibration wizard. Step 7.
Check that the measured value compared to
the reference instrument is within the accuracy
for the input. If not Click
Re-calibrate
button.
If correct click
Cancel
button to exit Calibration Wizard.
7.3.6 ENTER PASSWORD.
When saving for the first time a dialog will appear to enter password.
Default password from factory is
elcon
Note!
This dialog will only appear once.
BTS11 will keep the password in memory.
After successful have entered the password.
This dialog appears.
Date:
Todays date is displayed.
Signature:
Enter the Company/Your signature.
Click
OK
to save calibration or
Cancel
to abort.