Figure 7. Perpendicular filament alignment.
4. Replace and tighten the two retaining screws.
5. Switch the unit on and press 2nd/SET CAL (single/dual angle models only) or SET CAL (triple angle
models only).
6. Place the gauge on the calibration tile and hold down the green READ button. Use a small screwdriver
to adjust the internal calibration screw(s), located at the left-hand end of the bottom PCB (Figure 8);
determine which screw to adjust using the table below.
Gauge type
Angle
Adjust this screw
Single angle
20/45/60/75/85°
C
Dual angle
20°
C
Dual angle
60°
B
Triple angle
20°
C
Triple angle
60°
B
Triple angle
75/85°
A