R
17
1.
Enter system set-up and then press the SCROLL UP/DOWN key
until the display shows:
2.
Press the READ/SELECT key to enter this menu option. The display
will show:
3.
Use the SCROLL UP/DOWN key to adjust the value until it matches
the value shown on the calibration tile in use.
4.
When the displayed value matches that shown on the calibration tile,
press the READ/SELECT key to store the value. The gauge will
confirm this with the following message:
5.
The gauge will then return to the measurement mode and the gauge
must be re-calibrated before it is used for gloss measurements - see “Calibration” on page 14.
7.2 ZERO CALIBRATION - SWITCHING ON/OFF
Calibration between two points (high gloss and zero gloss) can be selected as an option.
To switch zero calibration on or off use the following procedure:
1.
Enter system set-up and then press the SCROLL UP/DOWN key
until the display shows:
2.
Press the READ/SELECT key to toggle zero calibration on/off.
3.
Press the C key to exit set-up mode.
SET STD CAL VAL.
60°
88.1
SET STD CAL VAL.
CAL Mode Active
SET CAL UP
60°
88.2
CAL DATA SAVED
Completed
Zero Calib. Off
Press Read Key
Zero Calib. On
Press Read Key