R
15
3.
Dual angle models only: In measurement mode, select the required angle, pressing the C key to cycle
through the angles 20 > 60 > 20/60. To calibrate two angles simultaneously, select the two angle
measurement mode ‘20/60’.
4.
Press and hold the C key to begin the calibration procedure. The
display will show:
5.
Press the READ/SELECT key. After a short delay during which a
‘Please Wait’ message appears, the display will show:
6.
Press
d
the READ/SELECT key. After a short delay during which a
‘Please Wait’ message appears, the display will show:
The gauge will then return to the measurement mode and is ready for use.
Note: The Elcometer 406L has an automatic calibration tile check routine. If the calibration tile is dirty or
damaged and the value does not conform to that stored in the memory the display will show the following
warning messages:
d.
If the READ/SELECT key is not pressed at this stage the calibration will be cancelled.
CALIBRATE ANGLE 60°.
Press Read Key
CALIBRATE
Sample High Ref
60° Angle
Completed
Poss. Cal. Error
Check Calib Tile
Accept new
Calibration?