Manuale di istruzioni/Scheda tecnica

La divulgazione dei dati contenuti in questa scheda è da ritenersi un servizio puramente informativo e non costituisce alcun vincolo da parte della Elcart in merito a prestazioni ed utilizzo del prodotto.

The divulgation of data contained on this technical sheet are exclusively for informational reasons and establish no link on behalf of Elcart regard to thr performances and the usa of the product.

La divulgacion de los datos contenidos en esta ficha son un servicio unicamente informativo y no constituyen ningun vinculo de parte de Elcart respecto a las prestaciones y uso del producto.

ART. 9/9450

PAGINA 9 DI 9

4.2.6 Transistor

RANGE

DESCRIPTION

TEST CONDITION

hFE

Display read approx. HFE value 

(0-1000) of transistor under test 

(all type)

Base Current approx. 

10μA, Vce approx. 2.8V

4.2.7 Inductance

RANGE

RISOLUZIONE

PRECISIONE

20mH

0.01mH

±(3.0% of rdg+5digits)

200mH

0.1mH

±(3.0% of rdg+5digits)

2H

1mH

±(3.0% of rdg+5digits)

20H

10mH

±(3.0% of rdg+5digits)

Overload protection:  

fuse (FF400mA/1000V)

4.2.8 Capacitance

RANGE

RESOLUTION

ACCURACY

20nF

10pF

±(4.0% of rdg+3digits)

200nF

0.1nF

±(4.0% of rdg+3digits)

2µF

1nF

±(4.0% of rdg+3digits)

200µF

100nF

±(5.0% of rdg+10digits)

Overload protection:   

fuse (FF400mA/1000V)

4.2.9 DC Current

RANGE

RESOLUTION

ACCURACY

200mA

0.1mA

±(1.5% of rdg+1digits)

10A

10mA

±(2.0% of rdg+5digits)

Overload protection:  

fuse (FF400mA/1000V) 

   

10A range fuse(FF10A/500V)

Max. input current:  

200mA dc or 200mA ac rms for mA range 

   

10A dc or 10A ac rms for 10A ranges

For measurements>5A: 
   

10 Seconds maximum ON to measure 1 minutes OFF 

   

Above 10A unspecified

4.2.10 AC Current 

RANGE

RESOLUTION

ACCURACY

200mA

0.1mA

±(1.8% of rdg+3digits)

10A

10mA

±(3.0% of rdg+7digits)

Overload protection:  

Resettable fuse (FF400mA/1000V) 

   

10A range fuse(FF10A/500V)

Max. input current:  

200mA dc or 200mA ac rms for mA range 

   

10A dc or 10A ac rms for 10A ranges

For measurements>5A: 
   

10 Seconds maximum ON to measure 1 minutes OFF 

   

Above 10A unspecified

Frequency Range:  

40Hz-400Hz

Response:  

Average, calibrated in rms of sine wave

5. MAINTENANCE

Do not attempt to repair or service your Meter unless you are qualified 

to do so and have the relevant calibration, performance test, and service 

information.

5.1 General Maintenance

WARNING

To avoid electrical shock or damage to the meter, 

do not get water inside the case. 

Remove the test leads and any input signals before 

opening the case

Periodically wipe the case with a damp cloth and mild detergent. 

Do not use abrasives or solvents.

Dirt or moisture in the terminals can affect readings.

To clean the terminals:

- Turn the meter off and remove all test leads.

- Shake out any dirt that may be in the terminals.

- Soak a new swab with a cleaning and oiling agent (such as WD-40).

- Work the swab around in each terminal. The oiling agent insulates the 

terminals from moisture-related contamination.

5.2 Battery and Fuse replacement

WARNING

To avoid false readings, which could lead to possible electric 

shock or personal injury, replace the battery as soon 

as the battery indicator (          ) appears. 

Use only fuses with the amperage,interrutp,voltage

And speed ratings specified

FF400mA/1000V Min interrupt rating 10000A

FF10A/500V  Min interrupt rating 10000A

To replace the battery of fuse:

- Turn the meter off. 

- Disconnect test leads and/or any connectors from the terminals.

- Use a screwdriver to unscrew the two screws secured on the battery    

 cover. 

- Take out the battery cover from the meter.

- Remove the used batteries or damaged fuse.

- Replace with the new 9V batteries (6F22) or new fuse.

- Rejoin the battery cover and secure by the two screws.

IMPORTED AND DISTRIBUTED BY:

ELCART DISTRIBUTION SPA

Via Michelangelo Buonarroti, 46

20093   COLOGNO MONZESE   (MI)

ITALY

www.elcart.com - [email protected]

Information for users:

The symbol on the equipment indicates that the waste must be “separately collected”.

Therefore, the user must carry (or have it carried) the waste to the separately collected 

waste centers set up by local governments, or deliver it to the dealer against purchase 

of a new equivalent-type equipment.

The separate waste collection and the subsequent processing, recovery and disposal 

operations favour the production of equipment with recycled materials and limit the 

negative effects on the environment and on health which may be possibly caused by 

the waste improper management.

The improper product disposal by the user causes the application of administrative 

sanctions according to the Art. 50 et. seq. of the Law Decree No. 22/1997.           

Made in China

Summary of Contents for 9/9420

Page 1: ...isurazioni effettuate nel l ambito domestico La categoria di misurazione II relativa alle misurazioni effettuate su circuiti collegati direttamente a una bassa tensione La categoria di misurazione I r...

Page 2: ...isplay 4 Pulsanti 5 Manopola 6 Terminali 2 2 Display LCD SIMBOLO SIGNIFICATO La batteria scarica Per evitari guasti e malfunzionamenti sostituite la batteria Indicatore letture negative Indicatore per...

Page 3: ...000m Temperatura di utilizzo 0 C 40 C 32 F 122 F 80 RH 10 C senza condensa Temperatura di conservazione 10 C 60 C 14 F 140 F 70 RH rimuovere la batteria Coefficiente di temperatura 0 1 precisione spec...

Page 4: ...NZIONE Non tentate di riparare lo strumento Solo persone qualificate possono effettuare riparazioni tecniche sul multimetro 5 1 Manutenzione generale ATTENZIONE Per evitare danneggiamenti non utilizza...

Page 5: ...easurement category I is for measurements performed on circuits not directly connected to MAINS When using this Multimeter the user must observe all normal safety rules concerning protection against t...

Page 6: ...e and ensure that it cannot be used until it has been checked out If the meter is not going to be used for a long time take out the battery and do not store the meter in high temperature or high humid...

Page 7: ...ircuit being measured and read the displayed value NOTE On 20M and 200M ranges the meter may take a few seconds to stabilize reading this is normal for high resistance measuring 3 2 3 Diode Test WARNI...

Page 8: ...ds and recover the measured circuit 4 TECHNICAL SPECIFICATIONS 4 1 GENERAL SPECIFICATIONS Environment conditions 1000V CAT II e 600V CAT III Pollution degree 2 Operating altitude 2000m Operating tempe...

Page 9: ...not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration performance test and service information 5 1 General Maintenance WARNING To avoid electr...

Reviews: