ART. 9/9450
PAGINA 8 DI 9
3.2.8 Current measurement
WARNING
To avoid damage to the Meter or injury if the fuse blows, never
attempt an in-circuit current measurement where the open-circuit
potential to earth is greater than 250V.
To avoid damage to the meter, check the meter's fuse
before proceeding.
Use the proper terminals, function, and range for your measurement.
Never place the test leads in parallel with a circuit or component
when the leads are plugged into the current terminals.
DC current ranges are:
DC: 200mA, 10A
AC: 200mA, 10A
To measure current:
1. Turn off the power of the measured circuit.
Discharge all the high voltage capacitors.
2. Set the rotary switch to the proper range.
3. Connect the black test lead to the COM terminal and the red to the
mA terminal for a maximum of 200mA.
For a maximum of 10A, move the red test lead to the 10A terminal.
4. Break the circuit path to be tested.
Connect the black test lead to the more negative side of the break;
connect the red test lead to the more positive side of the break.
(Reversing the leads will give a negative reading,but will not damage
the Meter)
5. Turn on the power of the measured circuit, and then read the display.
Be sure to note the measurement units at the right side of the display
(mA or A). When only the figure "1" displayed, it indicates overrange
situation and the higher range has to be selected.
6. Turn off the power of the measured circuit and discharge all the high
voltage capacitors.
Remove the test leads and recover the measured circuit.
4 TECHNICAL SPECIFICATIONS
4.1 GENERAL SPECIFICATIONS
- Environment conditions:
1000V CAT. II e 600V CAT. III
- Pollution degree:
2
- Operating altitude:
< 2000m
- Operating temperature:
0°C÷40°C, 32°F÷122°F
(<80% RH, <10°C non condensing)
- Storage temperature:
-10°C÷+60°C, 14°F÷140°F
(<70% RH, battery removed)
- Temperature coefficient: 0.1% (specified accuracy)/°C (<18°C o >28°C)
- MAX voltage between terminals and earth ground:
750VCA rms o 1000VCC
- Fuse Protection: mA: fuse (FF400mA/1000V) / A: fuse (FF10A/500V)
- Display: 3 1/2" digits LCD automatic indication of functions and symbols
- Over Range indication:
LCD will display "1".
- Low battery indication:
the " " is displayed when the battery is
under the proper operation range
- Polarity indication:
"-" displayed automatically
- Power source:
9V
- Battery type:
NEDA 1604, 6F22, or 006P
- Dimensions:
188(L)x92(W)x50(H) mm
- Weight:
380g. approx (battery included)
4.2 Measurement specifications
Accuracy is specified for one year after calibration, at operating tempera-
tures of 18°C to 28°C, with relative humidity at 0% to 75%.
Accuracy specifications take the form of:
±(% of R Number of Least Significant Digits)
4.2.1 DC Voltage
RANGE
RESOLUTION
ACCURACY
200mV
0.1mV
±(0.5% of rdg+1digits)
2V
1mV
±(0.5% of rdg+1digits)
20V
10mV
±(0.5% of rdg+1digits)
200V
100mV
±(0.5% of rdg+1digits)
1000V
1V
±(0.8% of rdg+2digits)
Input impedance:
10MΩ
Max input voltage:
250Vdc or ac rms for 200mV range
and 1000Vdc or 750V ac rms for other ranges
4.2.2 AC Voltage
RANGE
RESOLUTION
ACCURACY
2V
1mV
±(0.8% of rdg+3digits)
20V
10mV
±(0.8% of rdg+3digits)
200V
100mV
±(0.8% of rdg+3digits)
750V
1V
±(1.2% of rdg+3digits)
Input impedance:
10MΩ
Max. input voltage:
250Vdc or ac rms for 200mV range
and 1000Vdc or 750V ac rms for other ranges
Frequency Range:
40Hz-400Hz
Response:
Average, calibrated in rms of sine wave
4.2.3 Resistance
RANGE
RESOLUTION
ACCURACY
200Ω
0.1Ω
±(0.8% of rdg+3digits)
2KΩ
1Ω
±(0.8% of rdg+1digits)
20KΩ
10Ω
±(0.8% of rdg+1digits)
200KΩ
100Ω
±(0.8% of rdg+1digits)
2MΩ
1KΩ
±(0.8% of rdg+1digits)
20MΩ
10KΩ
±(1.0% of rdg+2digits)
200MΩ
0.1MΩ
±(5.0% of rdg+10digits)
Overload protection:
250V dc or 250Vac rms
Open Circuit Voltage:
less than 700mV
4.2.4 Diode
RANGE RESOLUTION
ACCURACY
1mV
Display read approx. forward voltage
of diode
Forward DC Current:
approx. 1mA
Reversed DC Voltage:
approx. 2.8V
Overload protection:
250Vdc or 250Vac rms
4.2.5 Audible continuity
RANGE
FUNCTION
The buzzer sound if the resistance of a circuit
under test is less than 30Ω
Test Current:
approx. 1mA
Reversed DC Voltage:
approx. 2.8V
Overload protection:
250Vdc or 250Vac rms