2.2.5 Alarm
Alarm is activated at:
• high temperature in the test chamber, + 1 °C from set value
• low temperature in the test chamber, - 1 °C from set value
• high temperature of the controller, + 5 °C from set value
1)
1)
This alarm will be reset automatically when the temperature is within tolerance from set value again.
Error code
Problem cause
Remedy
001–006
Oven cell is too warm.
The alarm will be reset when the alarm
Cell 1–6
has been acknowledged and the cell
has cooled down.
007–012
Temperature difference between
Wait until the temperature is within
Cell 1–6
set value and process value ± 1 °C.
range to acknowledge the alarm.
013
Low pressure from airpump.
Check the airpump filter and function.
014
Shinko Communication Timeout.
Contact support.
015
Low battery.
Replace backup battery in OPLC
(contact support).
016–021
Loop break controller #.
Broken temperatur sensor,
Cell 1–6
call for service.
022–027
Temperature difference
Acknowledge the alarm.
Cell 1–6
between set value of controller
When temperature is within range
and actual value + 5 °C.
the alarm will be reset. If the alarm
is not reset contact support.
Coloured lines on PV and SV display indicates alarm status.
• A red line above PV display = PV temperature d 1 °C from SV temperature.
• A light blue line below PV = PV temperatures differs - 1 °C from SV temperature.
• A red line above SV = SV d 5 °C from actual controller temperature.
• A green line below SV = PV temperature is within ± 1 °C from SV temperature.
SV temp. diff. alarm to general alarm collection list and flashing alarm button
will only be active for running channels (active count down meter).