6.3 Replacing the membrane and ring
The replacement of the membrane should be done when the user has
appreciated that it has lost quality, either by use or by damage thereof.
The steps for this replacement are as follows:
1. Using your thumb pressure to remove the ring and the corresponding
membrane. With no membrane, then clean the chestpiece of the
stethoscope
2. Align the new membrane with the chestpiece.
3. Check that the position of the ring corresponds to the graph on the
right, i.e., the thickest part should look inward and the finest one in
contact with the patient
4. Insert the ring laterally and guide with the fingers to make it enter
completely.
5. Press with your thumb to finish locating it carefully not to bend and
damage the membrane.
6. Check that the ring fixes completely the membrane and there are no
exposed parts.
1
3
2
4
5
6.4 Technical service
Modifications to the product are not permitted. It must be repaired by
authorised personnel. For repairs, send equipment to:
eKuore Technical Service
C/ Valencia 45 B. 46100
Burjasot, Valencia. España
Outside Spain, for information on the maintenance and repair of eKuore
Pro® Series, contact an authorized distributor or send an email to:
6.5 Replaceable System Parts
eKuore Pro Series’ replaceable parts are:
•
S/M/L Chestpiece (EP1000/EP1001/EP1002)
•
S/M/L Chestpiece membrane (EK1501/EK1503/EK1505)
•
S/M/L Chestpiece ring (EK1502/EK1504/EK1506)
•
USB cable (EP1500)
•
Earbuds (EP1202)
For more information, visit