7
www.eksmaoptics.com
Chapter 5
Q
UICK
S
TART
G
UIDE
Provide proper ground connection to the HV module. In order to prevent electrical breakdown,
ensure that the distance between the bottom of HV module and the surface below is more than
6 mm.
Connect the +24V DC power supply to the HV module. The DC power supply must be able to
provide the maximum current as per
2.2
T
ECHNICAL
S
PECIFICATIONS
.
Turn on +24V DC power supply.
Set the output voltage using the potentiometer (
#8
in
Figure 2
). Ways to control of the output
voltage:
1.
Measure the output high voltage directly if your control equipment is capable of
measuring high voltage.
2.
Alternatively, measure the voltage between points
#6
and
#7
in
Figure 2
. This voltage
converts to output voltage approximately as 1V~1 kV.
Turn off +24V DC power supply.
Connect the HV supply load. Check that the connection cables are firmly attached.
HV output can be turned off when pins
#6
and
#9
in
Figure 3
are interconnected.