7
Trouble shooting
Problem
Er1
: Measurement before the devise
is stabilized
Er2
: The device shows a rapid room
temperature change.
+
Er.3
: The room temperature is not
in the range 0°C to 40°C (50°F to
04°F)
Er:
Error 5-9 The system is not
functioning properly
Hi:
In Ear mode: The temperature
taken is higher than +4.°C (08°F)
In Scan mode: The temperature
taken is higher than +80°C (76°F)
Lo:
In Ear mode: The temperature
taken is lower than +34°C (93.°F)
In Scan mode: The temperature
taken is lower than -°C (-7.6°F)
Solution
Wait until the icons stop flashing
before measuring.
Allow the thermometer to rest in
a room for at least 30 minutes at
room temperature range 0°C to
40°C (50°F to 04°F).
Remove the battery, wait for
minute and replace the battery. If
the message reappears, contact
the retailer for service
Please select the target within the
specifications. If the malfunction
still exists, contact the nearest
retailer.
Change to new battery.
The device is not able to start
measurement.
GUARANTEE
EKS guarantees that all defects due to materials or manufacturing faults will be remedied
by replacement or repair for a period of years from the date of purchase. Please keep the
purchase receipt and guarantee card in a safe place. If you have any complaints, please
return the product to your dealer with the guarantee card and the receipt.
Guarantee Card
2 year guarantee
Bon de Garantie
2 ans de garantie
Garantiekarte
2 Jahre Garantie
Tagliando di garanzia
2 anni di garanzia
Tarjeta de garantía
2 años de garantía
Гарантийная карта
2 лет гарантии
Date of purchase - Date d'achat -
Kaufdatum - Data d’acquisto -
Fecha de compra - Дата покупки
Place of purchase - Point d'achat -
Ort des Einkaufs - Punto d'acquisto -
Establecimiento - Место покупки
Summary of Contents for 0620
Page 39: ...39 ...