EKO MS-093 Instruction Manual Download Page 11

EKO INSTRUMENTS CO., LTD. Sunshine Duration Meter MS-093 Instruction Manual Ver.4 

Pg. 10

2. Reset Box (Optional) 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

NOTE:    If Power Box is used, connect one end of the Reset Box Junction Cable to Power Box Terminal; if 

Power Box is not used, connect the Reset Box Junction Cable directly to DC Power Supply and Data Logger. 

(See [Table 5-3. Cable Colors and Outputs] for details.) 

 

 

3. Power Box (Optional) 

 
 
 
 
 
 
 
 
 
 
 
 

FG

Fuse 

Power Supply   

Connector 

Terminal 

Figure 5-4. MS-093 Power Box Parts Name 

Figure 5-3. Reset Box: Parts Name 

Terminal for

Connection Board 

(Connect Reset Box 

Junction Cable here)

Terminal for   
Sunshine Duration Meter 
(Connect Sensor Cable here)

Summary of Contents for MS-093

Page 1: ...INSTRUCTION MANUAL MS 093 Sunshine Duration Meter...

Page 2: ...1 WARNING CAUTION 5 4 Introduction 6 4 1 Main Functions 6 4 2 Package Contents 7 5 Getting Started 8 5 1 Sunshine Duration Meter Configuration 8 5 2 Parts Name and Descriptions 9 5 3 Installation 11 5...

Page 3: ...el 31 0 70 3050117 Fax 31 0 70 3840607 North America Region www eko usa com info eko usa com EKO INSTRUMENTS USA Inc 95 South Market Street Suite 300 San Jose CA 95113 USA Tel 1 408 977 7751 Fax 1 408...

Page 4: ...ollection points Disposing of this product correctly will help save valuable resources and prevent any potential negative effects on human health and the environment which could otherwise arise from i...

Page 5: ...elates is in conformity with the following harmonized standards of other normative documents Harmonized standards EN 61326 1997 A1 A2 Class A Emission EN 61326 1997 A1 A2 Immunity EN61000 4 2 EN61000...

Page 6: ...nexpected accidents Do not give an impact on exposed glass tube Do not hold the glass tube part when carrying Glass tube part may break and lead to serious injury and or accident by broken glass and o...

Page 7: ...ve the given threshold 120W m2 Also the pyroelectric sensor has flat characteristics thus it can correspond to the changes of solar spectral distribution changes by the climate 2 Mirror The mirror ins...

Page 8: ...rks MS 093 Main Unit 1 With Sensor Cable Standard cable length 10m Fixing Screws M5 x 2 Latitude Fixing Screw Hexagon Wrench 1 Set Includes 1 set of screw 1 hexagon wrench Reset Box 1 Optional Cable l...

Page 9: ...ctric circuit part contains control circuit components such as motor and CPLD 3 Reset Box Optional This unit monitors the movement of the mirror and when the mirror rotation stopped it resets the powe...

Page 10: ...93 Parts Name 1 Sensor Cable Connector 7 Latitude Scale Plate 13 Spirit Level 2 Drying Cartridge 8 Blower fan Unit Optional 14 Leveling Feet 3 Electric Circuit Part 9 Blower fan Unit Skirt 15 Blower f...

Page 11: ...Box is not used connect the Reset Box Junction Cable directly to DC Power Supply and Data Logger See Table 5 3 Cable Colors and Outputs for details 3 Power Box Optional FG Fuse Power Supply Connector...

Page 12: ...reflect solar radiation onto the sunshine duration meter The Sunshine Duration Meter is installed in North South direction Table 5 2 Orientation of Sunshine Duration Meter Installing Location Directio...

Page 13: ...ne See previous section for determining the North South direction 2 Attach the Sensor Stand on the Base Plate in advance Place the Base Plate with Sensor Stand on the mounting base so that the N side...

Page 14: ...Meter connector This connector is waterproof make sure to securely fasten the connector till the connector ring is completely closed 2 When Reset Box is used connect the Sunshine Duration Meter Senso...

Page 15: ...ine duration signals from MS 093 uses non voltage contact signal and it makes the contact for 1 second when sunshine is detected When taking measurement by connecting to a data logger MS 093 cannot be...

Page 16: ...eeds this threshold value it is determined as there is sunshine Moreover when the logic circuit recognizes the sunshine from the logic values which determine either sunshine or no sunshine the non vol...

Page 17: ...ent between sunshine not sunshine Check Horizontal Level Every Week Check and make sure the MS 093 base plate is in horizontal position and adjust the level as necessary Measurement error correspondin...

Page 18: ...ogger Mirror stopped rotating When Reset Box is not used Check the power supply make sure the power supply voltage is DC 12V If the power supply is in proper condition turn OFF the power once and put...

Page 19: ...o 450mA 30 to 40 Operation Temperature Range 20 to 40 Sunshine Duration Outputs Output Sunshine No Sunshine Non voltage contact output Pulse Width 1 0 05sec Voltage Resistance 60V Make contact for one...

Page 20: ...pe Power Consumption 11VA Size 200 W x 140 D x 80 H Weight Approximately 1 1kg 4 Blower Fan Unit Optional Table 8 4 Blower Fan Unit Specification Items Specifications Maximum Wind Flow Rate Approximat...

Page 21: ...oshi NJW 2012 PF 2 Power Cable Optional 3 Reset Box Junction Cable Optional 4 Pulse Output Cable Optional To Power Box Connector 2 5m To AC Plug Figure 8 2 Power Cable Figure 8 3 Reset Box Junction Ca...

Page 22: ...EKO INSTRUMENTS CO LTD Sunshine Duration Meter MS 093 Instruction Manual Ver 4 Pg 21 8 3 Dimensions 1 Sunshine Duration Meter MS 093 Figure 8 5 MS 093 Dimension Bottom of Sensor Stand...

Page 23: ...EKO INSTRUMENTS CO LTD Sunshine Duration Meter MS 093 Instruction Manual Ver 4 Pg 22 2 Reset Box Optional Figure 8 6 Reset Box Dimension...

Page 24: ...EKO INSTRUMENTS CO LTD Sunshine Duration Meter MS 093 Instruction Manual Ver 4 Pg 23 3 Power Box Optional Figure 8 7 Power Box Dimension...

Page 25: ...Cable Cable Length 1 5m Blower Fan Unit Cable Length 10m DC12V Base Plate Base Plate Fixing Bolts Nuts 1 set includes bolts and nuts 2pcs each Table 8 6 Consumable Parts List Parts Name Used Parts Qt...

Page 26: ...ncy of United nations who has authoritative role in standardization and control over international meteorological related activities World Radiation Reference WRR Radiometric reference instrument syst...

Page 27: ...eko co jp EKO North America 95 South Market Street Suite 300 San Jose CA 95113 USA P 1 408 977 7751 F 1 408 977 7741 info eko usa com www eko usa com EKO Europe Middle East Africa South America Lulofs...