8 Maintenance
Page 52 / 66
1. Only the above components can be autoclaved.
2. Before first use and after each use, sterilize the above components.
Reprocessing Instructions
Preparation at the Point of Use:
1. Disconnect the silicone protective cover and handheld bracket before
cleaning. Remove gross contaminations from the components with code water
(<40°C) immediately after use. Don’t use a fixating detergent or hot water
(>40°C) as this can cause the fixation of residuals which may influence the
result of the reprocessing process.
2. Store the instruments in a humid surrounding.
WARNING
Do not submerge the components or wipe them with any of the following
functional water (acidic electrolyzed water, strong alkaline solution, or ozone
water), medical agents (glutaral, etc.), or any other special types of water or
commercial cleaning liquids. Such liquids may result in metal corrosion and
adhesion of the residual medical agents to the components.
Transportation:
Safe storage and transportation to the reprocessing area to avoid any damage
and contamination to the environment.
Preparation for Decontamination:
The devices must be reprocessed in a disassembled state.
WARNING
Observe suitable personal protective measures.
Pre-Cleaning:
Do a manual pre-cleaning, until the components are visually clean. Submerge
the components in a cleaning solution. Clean the surfaces with a soft bristol
brush.
Cleaning:
Summary of Contents for NanoPix1
Page 1: ...Intraoral Digital Imaging Sensor USER MANUAL Changzhou Sifary Medical Technology Co Ltd...
Page 2: ...Version S04 IFU 6935013 S04 Issued 2020 12 19 Size 130mmx180mm...
Page 4: ...Table of Contents Page 4 66...
Page 7: ...2 Symbols Page 7 66 IPX8 Protection against liquid...
Page 18: ...4 Product Installing Page 18 66 2 Place unused sensors in the mounting bracket...
Page 26: ...5 Soft Introduction Page 26 66 Print Includes Paper Header and Footer Preference settings...