![Eighteeth HyperLight User Manual Download Page 31](http://html1.mh-extra.com/html/eighteeth/hyperlight/hyperlight_user-manual_2369382031.webp)
11 EMC Tables
Page
31
/
35
Conducted RF
IEC 61000-4-
6
Radiated RF
IEC 61000-4-
3
3 Vrms
150 kHz to 80
MHz
6Vrms in ISM
bands
between 0,15
MHz and 80
MHz
3 V/m
80 MHz to 2,7
GHz
(V1)=3Vrms
(E1)= 6Vrms in
ISM bands
(E1)=3V/m
Portable and mobile
communications equipment
should be separated from the
HyperLight by no less than the
distances calculated/listed below:
D=(3.5/V1)(Sqrt P)
150kHz to 80MHz
D=(3.5/E1)(Sqrt P)
80 to 800 MHz
D=(7/E1)(Sqrt P)
800 MHz to 2.5 GHz
where P is the max power in watts
and D is the recommended
separation distance in meters.
Field strengths from fixed
transmitters, as determined by an
electromagnetic site survey,
should be less than the
compliance levels (V1 and E1).
Interference may occur in the
vicinity of equipment containing a
transmitter.
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and D is
the recommended separation
distance in meters (m).
Guidance and Manufacturer
’
s Declaration
–
Electromagnetic Emissions
The HyperLight is intended for use in the electromagnetic environment specified below. The
customer or user of the HyperLight should ensure that it is used in such an environment.
Immunity
Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment
–
Guidance