User’s ManUal

Concentric Microdialysis Probe 

Model AZ-X-Y

•  All Eicom products are for animal research purpose only.
•  Keep the probe away from physical shock.
•  Inlet needle is longer than outlet needle.
•  Dura must be broken prior to insertion into the brain.

AZ-X-Y Probe

the  probe  membrane.  After  flushing  for  several  minutes  (it  is  not 
necessary to flush for an extended period of time) the membrane may 
“sweat” or appear wet this is normal, not due to damage or a clog.

2.  If no flow is observed via the outlet needle, soak the probe in distilled 

water  for  a  few  minutes  and  repeat  instructions  in  item  1.  After 
confirming the distilled water flow from the outlet, please remove the 
probe from plastic sheath and confirm no leaking.

3.  We  test  each  probe  for  defect/leaking  prior  to  shipping.  If  you  find 

a defective probe after following the above steps Eicom will replace 
the probe at no charge. We do not guarantee probes after insertion 
to animals.

Removing Air Bubble

1.  If there is an air bubble in the probe, please remove it. Because the 

air bubble will cover the active membrane site and keep the perfusate 
away. This will result in a lower recovery rate. The air bubble should be 
removed before study. If it is small, it will disappear in case of in vivo 
study but may not in vitro.

2.  Perfusate needs to be degassed before use.
3.  To remove the air bubble, please introduce a big air to the inlet and 

flush out. Do not connect outlet tubing at this time. The big air bubble 
will trap and remove a small air bubble.

Connecting Outlet

1.  Confirm neither leaking or air bubbles are present in the membrane. 

Set the flow rate on the syringe pump. A standard flow rate is 1-2 μl/
min. A high flow rate coupled with a long outlet tubing can result in 
damage of the probe membrane.

2.  Connect  Teflon  tubing  (model  JT-10)  or  other  tubing  to  the  outlet. 

A long tubing length (more than 150 cm) can cause leaking of the 
membrane or the liquid swivel due to the induced high pressure. If you 
use tubing longer than 150 cm, please avoid using flow rates higher 
than  2  μl/min.  Polyethylene  tubing  can  cause  confounding  peaks 
when analyzing samples using HPLC with electrochemical detection.

Cleaning Probes and Storage

Remove the probe from the system and flush with distilled water to wash 
out  salts.  Then  put  the  probe  in  the  distilled  water  and  place  at  4°C  for 
storage.  Keep  the  membrane  wet  to  prevent  it  from  shrinking.  Keep  the 
probe away from physical shock.

Cleaning Lines (Teflon, FEP tubing)

Bacteria generation and protein pollution of tubing can be a cause of low 
recovery rates, especially in the outlet tubing. In order to avoid this problem, 
wash the lines with 50 times diluted commercial kitchen bleach (chlorine) in 
pure water. Perfuse the solution into the lines (not the probe) and leave it for 
at least one hour, then thoroughly rinse with water.

Perfusion Solution

We are commend using a non-buffered aCSF solution because phosphate 
buffer solutions may precipitate calcium phosphate and can result in clogs.

Probe Check

1.  Inject distilled water using a 1 or 2 ml disposable (plastic) syringe. 

Connect  the  syringe  to  the  inlet  (longer  needle)  of  the  probe  using 
Biton tubing (JB-30). The probe may become clogged or damaged if 
you initially flush with a salt solution. We recommend using distilled 
water for the initial probe flush. Do not connect tubing to the outlet 
of the probe during flushing. Please confirm there are no leaks from 

AD12-0718

inner  0.105/0.040 mm PFS

outer  0.63/0.33 mm SUS 

inner  0.150/0.075 mm PFS

outer  0.63/0.33 mm SUS 

0.3/0.2 mm SUS

0.18/0.12 mm SUS

Inner Dead Volume

inlet:  0.05 

µ

L

outlet:  0.4 

µ

L

9.0 mm

12.0 mm

X + 5

PFS:  Polyimide-Resin 

         Coated Fused Silica

Membrane:  Artificial Cellulose
Cut off:  MW 50,000 
OD/ID:  0.22/0.20 mm

SUS (Stainless Steel) 316

Y

Model:  AZ-X-Y

X = (guide cannula length)

Y = (membrane length) 

Reviews: