Function
and
Safety
Check
and
Test
At
least
once
a
year,
the
biomedical
engineering
department
or
other
qualified
personnel
should
do
the
following
check
and
test
:
Check
of
the
connectors
and
mains
supply
cord
conditions.
Visual
check
of
the
mechanical
protections.
Check
of
the
protections
against
the
danger
due
to
liquid's
pouring,
dripping,
moisture,
liquid's
penetration,
cleanliness,
sterilization
and
disinfection.
Check
of
the
Equipment's
Data
on
the
Label.
Check
of
the
availability
of
the
Instruction's
Manual.
Check
the
functioning
of
the
H.F.
output
controls.
Check
the
uniformity
of
the
resistance
through
the
surface
of
the
patient
plate.
Test
the
earth
conductivity
resistance.
Test
the
earth
leakage
current.
Test
H.F.
leakage
current.
Control
of
the
neuromuscular
stimulation.
Control
of
the
accuracy
of
the
output
power.