
36
USER MANUAL
RETEVET™ PORTABLE ERG
T +49 7461 96 580 0 | F +49 7461 96 580 90 | [email protected] | www.eickemeyer.com
Cleaning and Disinfection
Warning!
Consult the cleaning agent and germicidal cleaner agent manufacturer instructions for their proper use and
germicidal effi cacy prior to their use.
Caution!
• Do not submerge the device in liquid or allow liquid to enter the interior of the device as this could damage the
electronics. Do not use automatic cleansing machines or sterilization.
• Follow these instructions and only use the cleaning or germicidal cleaner agent types listed or damage may
occur.
Cleaning the Ganzfeld
The white interior sphere that the patient looks into (the ganzfeld), should be cleaned when there is visible dust inside or
when the device fails to calibrate at the start of a test.
The ganzfeld can be cleaned with a compressed gas air duster to remove dust. A damp moistened with water or isopropyl
alcohol cloth may be used if compressed gas doesn’t work. Liquid cleaners may damage the LED lights and camera inside it.
Cleaning and Disinfecting the Exterior
Cleaning of the patient contacting parts of the device (eyecup and sensor strip lead) is recommended between patient uses.
The RETevet™ device is chemically compatible to wipes containing 70 % isopropyl alcohol and with wipes containing alkyl
dimethyl benzyl ammonium chloride. The use of other wipes may damage the device.
1. Remove all visible soil by wiping all exterior surfaces with a compatible wipe.
Ensure that all visible contamination has been removed.
2. Disinfect using a germicidal wipe labeled suitable for use on healthcare equipment and capable of low or intermediate
level disinfection, following the procedures and contact time recommended by the germicidal wipe manufacturer.
3. Inspect for any visible damage prior to use. Discontinue use if any abnormalities are found.
Replacement eyecups and sensor three plug electrode leads are available. See Purchasing Supplies and Accessories on page
45.
Sterilization
The RETevet™ device is not intended to be sterilized.
Biocompatibility
The patient-contact portion of the RETevet™ device complies with biocompatibility standard ISO 10993-1.