4
‐
20 mA OUTPUT
Zero and Span Adjustment
Zero and Span can be set anywhere within the range of the sensors listed. The 2
‐
wire 4
‐
20 mA
output is user
‐
selectable for either a standard square root output (0.5 power) or for Eastern
Instruments’ modi
fi
ed power output (0.455 power).
Accuracy: +/
‐
0.05% Full Scale (
‐
40
o
to 70
o
C)
Power Supply: Loop Power Required
Supply Voltage Range: 8.8 to 42.5 Vdc
Power Supply E
ff
ect: Negligible
Maximum load resistance is determined by the voltage level of the external power supply as
described in the graph (right):
Load
Resi
stan
ce
, OHM
Supply Voltage, vDC