GES 400 MR Series User Manual | 8100401-55
•
2015-07-22
23
•
Do not use the system if it has been damaged, until it has been
verified to be working correctly.
•
Only EGI support or authorized engineers may service this
equipment.
Hazardous mains voltage inside. Refer all servicing
to EGI Technical Support (Table P-3).
•
Do not remove any parts, while the GES is powered.
The GES
uses potentially dangerous line voltages, which are present
within some subsystem devices.
1.4.2
Patient Isolation
WARNING:
•
Connect all GES components, except those inside of the MR
imaging room
,
only to the power mains cord supplied with the
protective ground terminal.
The presence of ground protection
preserves the patient’s safety, as well as your own. Check that your
local electrical plant can guarantee an efficient grounding. The
grounding reliability of a GES can be ensured only when used with a
hospital-grade plug.
•
Connect all GES components, except those inside of the MR
imaging room
,
to the isolation transformer’s outlets.
Never
connect the computer or other system components directly to a
wall power socket, because the leakage currents in the
computer can present an electrical hazard to you and the
patient.
•
Ensure electrode isolation.
Prevent contact between the
conductive part of the electrodes and other conductive parts,
including the ground.
Summary of Contents for GES 400 MR
Page 4: ...4 GES 400 MR Series User Manual 8100401 55 2015 07 22...
Page 12: ...12 GES 400 MR Series User Manual 8100401 55 2015 07 22...
Page 52: ...52 GES 400 MR Series User Manual 8100401 55 2015 07 22...
Page 56: ...56 GES 400 MR Series User Manual 8100401 55 2015 07 22...
Page 64: ...64 GES 400 MR Series User Manual 8100401 55 2015 07 22...
Page 65: ...GES 400 MR Series User Manual 8100401 55 2015 07 22 65...