11
Testing for Continuity
To test for continuity, connect the Meter as below:
1. Insert the red test lead intoV terminal and the black test lead into the
COM
terminal.
2. Set the rotary switch to
.
3. Connect the test leads across with the object being measured.
4. The buzzer does not sound if the resistance of a circuit under test is
≤70Ω.
The tested circuit
resistance value simultaneously shows on the display and the unit is
Ω.
Note
When continuity testing has been completed, disconnect the connection between the testing
leads and the circuit under test, and remove the testing leads away from the input terminal of
the Meter.
E. FREQUENCY MEASUREMENT (see figure 7)
(figure 7)
Warning!
To avoid harms to you, never attempt to input voltages higher than 60V DC or V rms
in AC. When the voltage of tester frequency signal is higher than 30V rms, the reading
obtained may not be without the accuracy.
The frequency measurement range is 20kHz. To measure frequency, connect the Meter as
follows:
1. Insert the read test lead into the
V
Ω
terminal and the black test lead into the
COM
terminal.
2. Set the rotary switch to
20kHz
range.
3. Parallel connect the test leads across with the object being measured.
The measured value shows on the display.
Note
When Hz measurement has been completed, disconnect the connection between the testing
leads and the circuit under test, and remove the testing leads away from the imput terminals of
the Meter.