1000D -- Hookup
❶
Connect the air input hose to a plant air source. Set plant air supply
within 80 to 100 psi (5.5 to 6.9 bar). Where required, use an EFD
five-micron filter regulator #2000F755 (see Warranty).
❷
Attach the air input hose coupling to the dispenser. Pull back metal
ring to attach to dispenser.
❸
Plug in the polarized foot pedal connector.
❹
Check the voltage label on the input voltage selector cartridge. To
change the voltage, remove the voltage selector from the car-
tridge, rotate it and position the correct voltage to show through the
cartridge window. Replace the cartridge into the power cord
receptacle and insure that both sides snap securely into position.
Install the power cord.
❺
Attach the 10cc barrel pre-filled with blue, nontoxic test fluid
(included with the dispenser) to the 10cc adapter head.
❻
Take the 10cc adapter assembly (#5150 on the adapter head) and
insert the black, male quick-connect into the air output fitting on the
front panel and turn clockwise to lock. Place the barrel in the barrel
stand.
1000D
220
120
100
Spare fuse
Voltage value
Cartridge window
(check voltage indicated)