Page 32
A746‐03‐885A ‐ Maintenance
Figure 29
Assemble the fitting at the pump head
If the pump does not achieve the ultimate pressure
In case the diaphragms and valves have been replaced, a run‐in period of several hours is
required before the pump achieves its ultimate vacuum.
If pump does not achieve the ultimate total pressure:
Check hose connectors between pump heads and manifolds for leaks.
If necessary recheck pump chamber.
Table 9
Troubleshooting
Fault
Possible cause
Remedy
Pump fails to start or
stops immediately.
Supply voltage is missing or too
low?
Check or connect supply
voltage.
Only 24 V DC version:
Control signal for motor speed is
missing?
Check control signal.
Pressure in outlet pipeline too
high?
Remove blockade in line, open
valve.
Only pumps with dual‐voltage
motor:
Motor overloaded?
Manual reset is necessary.
Switch off the pump or isolate
the equipment from mains.
Wait approx. five minutes
before restarting the pump.
Identify cause of failure and
eliminate.
The motor fuse has blown
page 6). Identify and eliminate
cause of failure.
Summary of Contents for XDD1 115 V AC
Page 8: ...This page has been intentionally left blank Page iv A746 02 885A List of Tables ...
Page 16: ...Page 8 A746 03 885A Technical data Figure 3 Dimensional drawing XDD 1 dual voltage motor ...
Page 17: ...Page 9 A746 03 885A Technical data Figure 4 Dimensional drawing XDD 1 24 V DC ...
Page 44: ...This page has been intentionally left blank Page 36 A746 03 885A Spares and accessories ...
Page 45: ...This page has been intentionally left blank ...