Page 11
A746‐03‐885A ‐ Installation
Figure 6
Replacing the motor fuse (dual‐voltage motor)
1. Carefully remove the fuse holder at the side of the pump (refer to
item 5).
2. Replace the defective fuse with a 2.5 A type T and secure the fuse carrier back into its
holder.
XDD1 24 V dc
The pump has been factory set to a constant pumping speed when connected to a 24 V DC
(=/‐10%) supply. The pump can also be controlled using an external analogue 0‐10 V signal.
Use a suitable connector mating half (not supplied) to connect the electrical supplies and your
control equipment to the connector on the logic interface cable. When you make the electrical
connections to the XDD 1, refer to
on page 11 for full details of the logic interface
connections.
WARNING:
Ensure that the electrical installation of the XDD1 mains pump conforms with your local
and national safety requirements. It must be connected to a suitably fused and
protected electrical supply and suitable earth point. For electrical data refer to
Table 6
Logic interface connector pins
Pin number
Signal
Use
2
Control / monitor: 0 V signal
‐
4
XDD 1 identity
9
Speed (0‐10 V)
To control motor speed
8, 13, 14
Electrical supply: 0 V
Use all pins for connection!
1, 6, 11
Electrical supply 24 V
Use all pins for connection!
Summary of Contents for XDD1 115 V AC
Page 8: ...This page has been intentionally left blank Page iv A746 02 885A List of Tables ...
Page 16: ...Page 8 A746 03 885A Technical data Figure 3 Dimensional drawing XDD 1 dual voltage motor ...
Page 17: ...Page 9 A746 03 885A Technical data Figure 4 Dimensional drawing XDD 1 24 V DC ...
Page 44: ...This page has been intentionally left blank Page 36 A746 03 885A Spares and accessories ...
Page 45: ...This page has been intentionally left blank ...