Maintenance
The System Status word used in the example above was obtained with the pump at rest. Decoding the word provides
additional information about the state of the pump. Refer to
3
Vent valve closed
Vent valve energised
4
Start
Start command active
5
Serial enable
Serial enable active
6
Standby
Standby active
7
Half full speed
Above 50% full rotational speed
8
Parallel control mode
Exclusive control mode selection
9
Serial control mode
Exclusive control mode selection
10
Invalid Podule software
Podule internal software mismatch
11
Podule upload incomplete
Podule failed internal configuration and calibration
operation
12
Timer expired
Failure to reach or maintain half full speed within the
timer setting value
13
Hardware trip
Over-speed or Over-current trip activated
14
Thermistor error
Pump internal temperature measurement system
disconnected or damaged
15
Serial control mode interlock
Serial enable has become inactive following a serial
Start command.
Table 26 - Example decoding of system status words
Bit number
Status of bit
(in example)
Indication
0
0
The pump has not failed
1
1
The pump is at rest
2
0
Speed is below normal speed
3
0
The vent valve is open
4
0
There is no active Start command
5
1
Serial enable is active
6
0
Standby is not active
7
0
Speed is below 50% of full rotational speed
8
0
The pump is not in parallel control mode
9
0
The pump is not serial control mode
10
0
There is no Podule internal software mismatch
11
0
Podule passed internal configuration and calibration operation
12
0
The timer has not timed out
13
0
Over-speed and Over-current trip not activated
14
0
Pump internal temperature measurement system is fine
15
0
Serial enable has not become inactive during serial control
Table 25 - Status flags (continued)
Bit number
Status flag
Active flags means