Installation
A652–07–880 Issue D
3.9
Fit the inlet catchpot
1. Ensure that the pump inlet (
, item 19) is clean and that the centring ring and filter assembly (supplied
with the pump) is in place on the inlet.
2. Place the inlet catchpot (24) on the pump inlet so that the drain plug (23) points towards the front of the pump
and the arrow on the upper half of the body points towards the pump inlet, as shown on
.
3. Secure the inlet catchpot to the pump using a clamping ring (20).
4. Secure the hose adaptor (1) to the inlet (25) of the inlet catchpot using an O-ring and centring ring (27) and
clamping ring (26).
3.10
Electrical installation
3.10.1
Check and configure the motor
CAUTION
Ensure that the motor is correctly configured for the electrical supply. If the motor is configured for 110 V
operation and the pump is operated with a 240 V supply, the motor will be damaged.
Ensure that the voltage shown on the voltage selector switch (
, item 3) in the motor cover corresponds with
the local electrical supply voltage. If it does not, change the configuration of the pump motor to match the local
electrical supply voltage; use the following procedure.
Use the following procedure to change the configuration of the pump motor:
1. Undo the two retaining screws (6) securing the voltage selector switch cover (5).
2. Remove the voltage selector switch cover (5) and toggle the voltage selector switch (3) into the alternate
position.
3. Invert the voltage selector switch cover (5) and refit over the voltage selector switch (3).
4. Refit the two retaining screws (6).