Operation
D397-50-880 Issue C
Input bytes (to Master)
No Alerts currently exist for these software modules so the 2nd and 3rd bytes will always be zero.
Associated Parameter options
None.
4.8.13
Software module 909 - Turbo Cycle Time
This module can be used with any Turbo pump that is compatible with a TIC unit.
General Description
Monitors turbo pump running time, since it was last started.
Output bytes (from Master)
None.
Input bytes (to Master)
No Alerts currently exist for these software modules so the 4th and 5th bytes will always be zero.
Associated Parameter options
None.
4.8.14
Software module 910 - Backing pump control
This module can be used with any TIC that includes pump control.
General Description
Controls Backing Pump.
Output bytes (from Master)
Input bytes (to Master)
No Alerts currently exist for these software modules so the 2nd and 3rd bytes will always be zero.
Associated Parameter options
None.
1st
Whether Standby speed mode is set: 0 =No; 4 = Yes.
2nd
Alert - refer to table in Appendix A2.
3rd
Priority - refer to table in Appendix A3.
1st
First two bytes form a 16-bit number which is the turbo
running time in hours since the pump was started.
2nd
3rd
State - refer to table in Appendix A1.
4th
Alert - refer to table in Appendix A2.
5th
Priority - refer to table in Appendix A3.
1st
Backing pump control (1 = run; 0= stop).
1st
State - refer to table in Appendix A1.
2nd
Alert - refer to table in Appendix A2.
3rd
Priority - refer to table in Appendix A3.