Page 10
W602100880_A - Technical data
WARNING: HIGH VOLTAGE
Risk of electrical shock. The gauge can only be connected to power supplies, instruments
or control devices that comply with the requirements of a grounded Protective Extra-Low
Voltage (PELV) and Limited Power Source (LPS), Class 2. The connection to the gauge has
to be fused.
Table 5
Electrical
Supply voltage
at the gauge
+14 to +30 V d.c.
Class 2/LPS
ripple
1 V
pp
Current consumption
< 500 mA (maximum starting current)
Power consumption (supply voltage
dependent)
1 W
Fuse required
*
*
The gauge is protected against reverse polarity of the supply voltage.
1 AT (slow), automatic reset (Polyfuse)
Electrical connection
15-pin D-Sub, male
Sensor cable
without switching functions
5-pin plus shielding
with switching functions
9-pin plus shielding
Cable length
§
§
For longer cables, larger conductor cross-sections are required (R
cable
1.0
100 m (0.14 mm² conductor)
Grounding concept
Vacuum flange - signal common
See
Supply common - signal common
Conducted separately for differential
measurement (10
)
Table 6
Materials exposed to vacuum
Materials exposed to vacuum
Flange, tube
Stainless steel AISI 316L
Sensor and diaphragm
Ceramics (Al
2
O
3
99.5%)
Internal volume
3.6 cm
3
Admissible pressure (absolute)
200 Torr/mbar (F.S.)
4.0 bar (400 kPa)
1 to 100 Torr/mbar (F.S.)
2.6 bar (260 kPa)
0.1 Torr/mbar (F.S.)
1.3 bar (130 kPa)
Bursting pressure (absolute)
5.0 bar (500 kPa)
Summary of Contents for Barocel 7025
Page 1: ...W602100880_A Original Instructions Barocel 7025 Gauge INSTRUCTION MANUAL edwardsvacuum com...
Page 6: ...This page has been intentionally left blank Page 4 W602100880_A Contents...
Page 29: ...This page has been intentionally left blank...
Page 30: ...This page has been intentionally left blank...
Page 31: ......
Page 32: ...edwardsvacuum com...