Edwards APGX-H Instruction Manual Download Page 22

D023-92-880 Issue F

Page 16

© Edwards Limited 2018. All rights reserved.

Edwards and the Edwards logo are trademarks of Edwards Limited.

Maintenance

If the APGX-H is connected to an Edwards AGC, TIC or AGD, the display will indicate the corresponding pressure.

5.1.4

Reset factory calibration values

Note:

Performing a factory values reset will overwrite any existing calibration data stored in the tube.

The use of factory calibration reset will not correct a filament that has drifted with time or due to 
contamination.

Perform the following procedure in order to restore the calibration values when the gauge was initially 
manufactured:

1. Disconnect the power supply to the gauge.

2. With the power disconnected, press and hold the '

CAL

' switch.

3. Switch on the power to the gauge with the '

CAL

' switch  still depressed and hold the switch down until the status 

LED flashes red. This takes approximately six seconds. The factory calibration values will now be restored.

5.2

Clean the filter

CAUTION

You must remove the filter to clean it.  Do not clean the filter in position in the body tube as permanent damage 
to the filament can occur.

Refer to 

Figure 1

 and use the following procedure to clean the filter. The filter can only be replaced on NW flange 

versions of the gauge.

1. Switch off the electrical supply, ensure that the vacuum system is at atmospheric pressure and remove the 

APGX-H from the vacuum system.

2. Use circlip pliers to remove the filter-retaining circlip (3); ensure that you do not damage sealing surface on the 

flange or the inside of the body tube. Remove the filter (5).

3. Inspect the filter. If the filter is damaged or cannot be cleaned, replace the filter; a filter replacement kit is 

available as a spare (refer to 

Section 7

).

4. Use a suitable cleaning solution to degrease the filter, then wash the filter repeatedly in clean water.

5. Wash the filter with a suitable solvent to remove all traces of water and then thoroughly dry the filter.

6. Refit the filter into the body tube and refit the circlip.

5.3

Replacing the tube

Loosen the two screws and detach the tube from the electronics housing. Plug the replacement tube into the podule 
and secure the tube retaining screws (item 1 in 

Figure 6

). To ensure a good electrical connection between the tube 

and electronics housing, the securing screws should be tightened. The new tube contains calibration information 
preset in the factory however for optimum performance and accuracy it is recommended that the vacuum and 
atmosphere adjustment described in 

Section 5.1

 is carried out before use.

Summary of Contents for APGX-H

Page 1: ...2 880 Issue F Original Instructions Instruction Manual APGX H Description Item Number APGX H NW16 AL D023 91 000 APGX H NW25 ST ST D023 92 000 APGX H NW16 ST ST D023 95 000 APGX H 1 8 NPT ST ST D023 9...

Page 2: ...vant requirements of harmonised standards EN 61010 1 2010 Safety requirements for electrical equipment for measurement control and laboratory use General requirements EN 61326 2 3 2013 Class B Emissio...

Page 3: ...hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement in GB T 26572 X GB T26572 X Indicates that the hazardous substance contained in at least...

Page 4: ...This page has been intentionally left blank...

Page 5: ...GC TIC or AGD display 8 3 3 2 Connect to your own supply and control equipment 8 4 Operation 11 4 1 Pressure measurement 11 4 2 Set point adjustment 11 4 3 Error monitoring 12 5 Maintenance 15 5 1 Atm...

Page 6: ...trical Connection 9 4 Pressure Voltage Characteristics of the APGX H 11 5 Adjust the APGX H 13 6 Replace the APGX H Tube 17 Tables Table Page 1 Pins on the APGX H Electrical Connector Socket 8 2 Error...

Page 7: ...can measure the full range of pressures in any orientation The tube is detachable from the electronics housing and is available as a replacement part The APGX H requires a 14 5 to 30 V d c power supp...

Page 8: ...lled SET TRIP allows you to change the operating pressure of the set point device The set point output device is an open collector transistor If you use an Edwards AGC TIC or AGD the APGX H set point...

Page 9: ...e Edwards logo are trademarks of Edwards Limited Introduction D023 92 880 Issue F Figure 1 General View of the APGX H NW16 APGX H Shown 1 Cable connector plug 2 APGX H connector socket 3 Filter retain...

Page 10: ...D023 92 880 Issue F Page 4 Edwards Limited 2018 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited This page has been intentionally left blank...

Page 11: ...11 cm3 Enclosure rating IP40 Ambient temperature Operation 5 to 60 C Storage 30 to 70 C Ambient humidity operation 80 RH up to 31 C decreasing linearly to 50 at 40 C and above Operating conditions 200...

Page 12: ...s exposed to vacuum Figure 2 Dimensions mm NW16 APGX H Shown Hysteresis 500 mV Max external load rating 30 V d c 100 mA Max on voltage 1 3 V at 100 mA Gauge identification resistance 39 k Filament Tun...

Page 13: ...rated in nitrogen whilst vertical For correct pressure indication in your chosen gauge orientation the gauge should be recalibrated at atmospheric pressure Edwards recommends mounting the gauge tube v...

Page 14: ...s controllers have adequate immunity Table 1 Pins on the APGX H Electrical Connector Socket Note Do not connect the electrical supply ground pin 2 to the signal ground pin 5 If you do the APGX H outpu...

Page 15: ...ure falls to the set point value the transistor output changes to on closed or low impedance and the set point LED switches on amber The set point has a fixed hysteresis of 500 mV the set point output...

Page 16: ...D023 92 880 Issue F Page 10 Edwards Limited 2018 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited This page has been intentionally left blank...

Page 17: ...ressure If the APGX H is connected to a voltmeter convert the measured voltage to the corresponding pressure using the following equation P 10 V 6 mbar P 10 V 6 125 Torr P 10 V 4 Pascal where V is the...

Page 18: ...voltage which corresponds to a given pressure The APGX H has an error monitoring facility which ensures that the set point output signal is off For approximately 2 seconds immediately after the APGX H...

Page 19: ...18 All rights reserved Page 13 Edwards and the Edwards logo are trademarks of Edwards Limited Operation D023 92 880 Issue F Figure 5 Adjust the APGX H 1 SET TRIP switch 2 Set point LED 3 CAL switch 4...

Page 20: ...D023 92 880 Issue F Page 14 Edwards Limited 2018 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited This page has been intentionally left blank...

Page 21: ...allow the gauge to operate for at least 10 minutes 2 Press the CAL switch The status LED will flash red and the voltage output will read approximately 1 9 V If the APGX H is connected to an Edwards A...

Page 22: ...ure to clean the filter The filter can only be replaced on NW flange versions of the gauge 1 Switch off the electrical supply ensure that the vacuum system is at atmospheric pressure and remove the AP...

Page 23: ...ction 5 1 If this fails to solve the problem you will need to replace the tube Refer to Section 7 2 Gauge indicates calibration error Check the electrical connections and ensure pin 7 is not connected...

Page 24: ...D023 92 880 Issue F Page 18 Edwards Limited 2018 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited This page has been intentionally left blank...

Page 25: ...nvironmental requirements Alternatively you may be able to recycle the APGX H and or cables contact Edwards or your supplier for advice also see below The APGX H and associated cables are within the s...

Page 26: ...D023 92 880 Issue F Page 20 Edwards Limited 2018 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited This page has been intentionally left blank...

Page 27: ...odel and Item Number of your equipment Serial number if any Item Number and description of part 7 2 Spares The Filter Replacement Kit contains 5 filter retaining circlips and 5 filters 7 3 Accessories...

Page 28: ...D023 92 880 Issue F Page 22 Edwards Limited 2018 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited This page has been intentionally left blank...

Page 29: ...This page has been intentionally left blank...

Page 30: ...This page has been intentionally left blank...

Page 31: ...of Vacuum equipment and Components Procedure HS1 If you are returning a vacuum pump note the following If a pump is configured to suit the application make a record of the configuration before return...

Page 32: ...P800 80 000 Issue T edwardsvacuum com...

Reviews: