© Edwards Limited 2018. All rights reserved.
Page 5
Edwards and the Edwards logo are trademarks of Edwards Limited.
Technical Data
D023-92-880 Issue F
2
Technical Data
2.1
Mechanical data
2.2
Performance, operating and storage conditions
2.3
Electrical data
Note:
The electrical power consumption of the APGX-H is reduced when the system is at vacuum. The relationship
between the output signal voltage and pressure is described in
Dimensions
Refer to
Mass
APGX-H NW16 ALI
110 g
APGX-H NW16 ST/ST
160 g
APGX-H NW25 ST/ST
170 g
APGX-H
1
/
8
NPT ST/ST
150 g
Volume of gauge tube
11 cm
3
Enclosure rating
IP40
Ambient temperature
Operation
+5 to +60°C
Storage
-30 to +70°C
Ambient humidity (operation)
80% RH up to 31°C decreasing linearly to
50% at 40°C and above
Operating conditions
2000 m (indoor use only)
Maximum internal pressure
10 bar absolute (9 bar gauge)
Pressure range
1333 to 3 x 10
-4
mbar (1000 to 2.3 x 10
-4
Torr)
Filament operating temperature (above ambient)
100°C
Pollution degree
2
Electrical supply
Type
NEC Class 2 or Limited Power or Limited Energy
Voltage
+14.5 V to +30 V d.c.
Maximum power consumption
1.5 W
Start-up current
120 mA
Electrical connector
FCC68 / RJ45 type, 8-way
Pressure output signal
Range
1.9 V to 9.125 V d.c.
Error range
output < 1.8 V d.c. or output > 9.2 V d.c
Min load impedance
10 k
Max output current
1 mA
Set-point
Adjustment range
1.8 V to 9.3 V