© Edwards Limited 2015. All rights reserved.
Page 3
Edwards and the Edwards logo are trademarks of Edwards Limited.
Technical data
D146-71-880 Issue E
2
Technical data
2.1
Mechanical data
2.2
Performance, operating and storage conditions
2.3
Electrical data
Dimensions
See
Mass
840 g
Volume of gauge tube
26 cm
2
Enclosure rating
IP40
Ambient temperature
Operation
5 to 60°C
Storage
0 to 70°C
Ambient humidity (operation)
10 to 90% (non condensing)
Maximum operating altitude
2000 m
Maximum internal pressure
10 bar absolute (9 bar gauge)
Pressure measurement range
1 x 10
-8
to 1 x 10
-2
mbar (AIM-SI1/AIM-SLI1)
1 x 10
-9
to 1 x 10
-2
mbar (AIM-XI1/AIM-XLI1)
Pollution category
EN61010 Category 2
Voltage
+13.5 to +36 V d.c.
Max voltage ripple
1 V peak to peak
Max source resistance
50
Max power consumption
2 W
Electrical connector
FCC68/RJ45 type, 8-way
Pressure output signal
Range
2
output
10 V d.c.
Error range
output < 2 V d.c. or output > 10 V d.c.
Impedance
0.1
Min load impedance
10 k
Max current source
1 mA
Gauge enable
Control sense
Active low
Active level
< 1.3 V
Control impedance
12 k
pull up to positive supply
Set-point output external load rating
40 V d.c., 100 mA max
Set-point trip level
Range
1.8 to 10.0 V
Minimum load impedance
200 k
Hysteresis
500 mV