Instrument Test Procedure 

• 
• 

• 

• 
• 

• 

Ensure that the shorting plug is connected to the BNC pH in­

put at the back of the meter. Switch on using the switch on the 

back panel. 

Press the Mode key to select pH mode and clear any calibra­

tion data by pressing the Clear key for 5 seconds. 

The displayed reading should read 7.00 ±0.02pH 

Switch to m V mode by pressing the mode key. Display should 

read 0.0 ±0.2mV 

Switch to °C key by pressing the mode key. Ensure that the 

display has a value in the range 0.0 to 19.0

°

C and is adjustable. 

i.e. responds to the use of the -G iJ- keys. Adjust to 20.0

°

C

The meter is ready for calibration

pH/ION INPVT 

FIEf 

'-

'v 

kARL FISCHER 

Iv  v·I 

D.C. INfflT  9V  O.SW 

·-

..... -�''"".,..,,,.. ....... ,...,""' 

TEMP 

RS 232 

Figure 2. Rear view of RE357T:x 

A Note on Electrodes 

pH, Ion Selective and Redox electrodes may all be used with the 

RE357. These may be combination or half-cell types. Combination 

or sensing half-cells should have a BNC terminal and be connected 

to the socket marked pH/ION input at the back of the meter. Refer­

ence half-cells should have a 4mm bunched terminal and be con­

nected to the socket marked REF. Always refer to electrode in­

structions before use. See also, Appendix I. 

Summary of Contents for RE357

Page 1: ...test Section 2 6 pH Calibration Measurement Section 3 8 mV Calibration Measurement Section 4 8 Use ofthe Recorder Output Section 5 9 Use with a Printer or Computer Section 6 11 Troubleshooting Sectio...

Page 2: ...ct ION Ltd or your agent When you receive the shipment inspect the container for any signs ofdamage Note any evidence ofrough handling in transit Imme diately report any damage to the agent Note The c...

Page 3: ...into the power socket at the back ofthe meter then connect to the AC supply Battery Installation Approximately 24 hours ofcontinuous use is afforded by the 9V batte1y The BAT flag appears on the disp...

Page 4: ...ge 0 0 to 19 0 C and is adjustable i e responds to the use ofthe G iJ keys Adjust to 20 0 C The meter is ready for calibration pH ION INPVT FIEf v W kARL FISCHER Iv v I D C INfflT 9V O SW TEMP RS 232...

Page 5: ...Alter displayed reading Enters displayed value as calibration data Clears Calibration data whenpressed or 5 s Selects mV Mode SelectspH Mode Transmits data toprinter fx Version On RE357 labelled Hold...

Page 6: ...he Enter key The fist calibration point will now be entered automati cally at the correct value for the temperature ofmeasure ment The Cal flag will be displayed 0 Rinse blot and place the pH electrod...

Page 7: ...e CAL flag will flash Press Enter the CAL flag will stop flashing Manual Temperature Compensation Ifthe use ofa temperature probe is inappropriate e g small sample size then manual temperature compens...

Page 8: ...Follow the procedure for Absolute mV from Oto 0 8 Immerse electrode s in the standard or blank solution Wait for a stable reading and press Enter The display will automatically zero and the CAL flag...

Page 9: ...intout ofthe displayed reading and temperature only 0 To obtain a printout ofother parameters for the same sam ple press the Mode key and then the Send key To print a new identifier press and hold dow...

Page 10: ...F pH 7 06CR LF T 15 8 CCR LF mV 0 2mV CRLF 2CRLF Second calibration point _________________ PR Send Probe Status Data Command PR Sl 95 3 NERNSTCRLF E0 5 2mVCR LF RD Send Current Readings Command RD pH...

Page 11: ...c readings or Electrodes disconnected display reads on left hand Electrodes not immersed in solu side tion Reference electrode not filled Referencejunction dry Drifting readings Inconsistent or lack o...

Page 12: ...or buffers Electrodes out of solution In the event ofa malfunction it is important to pinpoint the prob lem to either the meter or the cell Ifa spare cell is available sub stitute it for the one in us...

Page 13: ...Relative mV offset Auto Range change at C Range Resolution C Accuracy Recorder Output Karl Fischer Output Display Power Instrument Size Instrument Weight 0 00 to 15 00 0 02pH 0 100 C 2 points 4 7 or...

Page 14: ...H storage solution in the protective cap and place over the glass bulb Never store the elech ode in distilled or de ionised water Never allow the elech ode to dry out pH electrode storage solution is...

Page 15: ...mperature coefficient information Table of Temperature Coefficients For EDT Buffers oc pH4 pH 7 pH 10 10 3 99 7 07 10 I8 15 4 00 7 04 10 14 20 4 00 7 02 10 06 25 4 00 7 00 10 00 30 4 00 6 99 9 95 35 4...

Reviews: