T3D User  Manual

030705

Press and hold the 

function

 switch until the segment next to 

speed

 lights. Rotate the speed paddle

wheel and the 0 and 1 segments will light indicating the state of the output signal.  Spin the
paddle wheel rapidly and the bar graph lights progressively.

Battery test

When the T3D is first turned on,  all bar graph segments flash briefly to test the segments,
followed by one segment.  The position of this segment displays the battery voltage.  If it
illuminates at 6 or lower, change the battery.  Replace with an alkaline 9 volt battery similar to
EVER READY #522 or equivalent.

Calibration

To insure the best accuracy the T3D should be re-calibrated if the ambient temperature has
changed by 30 degrees Fahrenheit or more relative to the last calibration.  To calibrate the unit
the power must be off and no transducer connected.  With the unit in the power off condition, 
press and hold the 

function

 button then press the 

power

 button.  Do not release the 

function

button for three seconds.  Successful calibration is indicated when all segments go out. 

Multi-function connector wiring:

pin

function

1

Speed sensor signal.

2

Speed 5 VDC power.

3

Transducer element.

4-5, 7-8

Ground connections.

6

Temperature sensor.

Pin 1 is at the five o’

clock position as viewed from the

front panel and the pins are counted in a
counterclockwise direction.

Optional Accessories

Test Cable  (RCA to Clip Leads)   P/N  004-TESTC
8 pin Jack  (for making Transducer Adapters)   P/N  520-001 

Warranty Information

Unit will be repaired free of charge for one year from date of purchase providing there is no
water damage or other evidence of improper use or handling.  Purchaser must ship unit prepaid
to address below;  EDI will pay the return freight.

For repair ship to:

       Electronic Devices, Inc.
       3140 Bunch Walnuts Road
       Chesapeake, VA.   23322
       ATTN:  Service Department
       
Please enclose a note describing the problem.