Edgetech PPM1 Operator'S Manual Download Page 13

13 

NOTE:  IT IS GOOD PRACTICE TO INSTALL A SHUTOFF VALVE ON EACH 
SIDE OF THE MEASUREMENT CELL, TO KEEP IT DRY WHEN OPENING 
THE SAMPLING SYSTEM. 
 

5.2 MEASURING DRY GASES 
 

5.2.1  MEASUREMENT TIME 

Measuring extremely dry gases is an art.  Time and patience are involved.  If you 
are in a hurry, you may not obtain accurate results.  As discussed above, all 
materials are hygroscopic.  That is, they have pores that retain water molecules.  
When the level of moisture in a gas is decreased, for example, it may take a very 
long time for all of the materials in the entire sampling system to outgas, or give 
off all the excess water molecules that are trapped in the pores of every inside 
surface of the system.  Depending upon your unique conditions, it could take a 
number of hours for complete stabilization of the moisture level in your gas.  Be 
sure you wait long enough before taking readings.   
 

5.2.2  MATERIALS 

As discussed in the Installation chapter of this manual, selection of materials is 
extremely critical for proper dry gas sampling.   For dry gas measurements, do 
not use PVC or other porous tubing.   

Some plastic tubing such as Impolene™ 

may be used for gases in the higher ppm ranges.  For measuring gases in the 
lower ppm ranges, only stainless should be used.  
 
If readings do not seem to stabilize after waiting a reasonable amount of time, 
check to see if you have a leak from outside the system where the vapor 
pressure is higher, or have excessively hygroscopic tubing, or possibly your 
stainless tubing is contaminated and requires cleaning. 
 

5.3 OPERATING THE INSTRUMENTS 
 

5.3.1  THE PPM1 

There are no controls to operate on the PPM1.  The operator simply monitors the 
digital display, or allows the analog and digital outputs to communicate with the 
data acquisition system.  When you first turn on the system, a short self-check 
period occurs on the display, and then it quickly starts to show the two moisture 
parameters that it is programmed to display.  These are changeable in the field, 
as shown in the following section on the Serial Port.  The digital displays, as well 
as the electrical outputs, are constantly updated. 
 

5.3.2  THE  PPM2

 

The PPM2 is a portable version of the PPM1.  It is necessary to charge the built-
in batteries before using it, and then to recharge them periodically.  The charger 
is supplied with the system. 
 
 

Summary of Contents for PPM1

Page 1: ...nalyzer OPERATORS MANUAL 19 Brigham Street Unit 8 Marlborough MA USA 01752 Tel 508 263 5900 800 276 3729 Fax 508 486 9348 E mail h2o edgetechinstruments com www edgetechinstruments com Tel 508 263 590...

Page 2: ...required 7 Wire and plug in the RS 232 connector if required 8 Replace the outer cover Do not over tighten the screws 9 Connect Sensor to dry gas source Allow system time to dry out 10 Turn on the Po...

Page 3: ...Wiring 10 4 3 1 Sensor Connections 10 4 3 2 Power Connections 10 4 3 3 Analog Outputs 10 4 3 4 Alarm Relays 10 4 3 5 RS 232 Serial Port 10 5 0 Operation 5 1 Startup 12 5 1 1 Dry Down Phase 1 12 5 1 2...

Page 4: ...s 23 7 0 Specifications 24 8 0 Appendix 26 8 1 PPM1 Mounting Dimensions 26 8 2 Sensor Dimensions 27 8 3 Warranty 28 8 4 Notes 29 8 5 Basic Humidity Definitions 30 LIST OF ILLUSTRATIONS 3 1 Phosphorous...

Page 5: ...want you to be entirely satisfied with your instrument The information contained in this manual will get you started It tells you what you need to get your equipment up and running and introduces its...

Page 6: ...mable Alarm Relays and also provides information to the Digital Display and Analog Outputs A microprocessor controls all data processing and input and output functions of the Instrument ensuring a hig...

Page 7: ...c cell the rate of gas flow must be carefully Figure 3 1 The Phosphorous Pentoxide controlled by the user for P2O5 Moisture Sensor accurate measurements this cell is independent of gas flow rate Altho...

Page 8: ...nstrument is programmed for the following standard conditions These may be field modified by the user by using the RS 232 serial port See the Operation section for complete programming information Dig...

Page 9: ...holes as the cover screws They must have a head diameter that is small enough to fit inside the clearance holes but large enough to press against the collar located at the bottom of the holes No 8 pan...

Page 10: ...nnected to the terminal marked CH GND 4 3 2 POWER CONNECTIONS Power is supplied by a user furnished unregulated DC power supply Voltage range may be between 18 and 30 VDC at 100 mA maximum Connect the...

Page 11: ...11 Figure 4 1 PPS1 PPS2 Wiring Connections Digital Display 4 20 mA Output 1 4 20 mA Output 2 Sensor Cable Connections DC Power Connections Alarm Relay K2 Alarm Relay K1 RS 232 Type DB 9 Connector...

Page 12: ...teps are taken to remove any moisture in advance by following the PHASE 1 dry down procedure much of the moisture desorbed from the surfaces of the sampling system will enter the electrochemical cell...

Page 13: ...use PVC or other porous tubing Some plastic tubing such as Impolene may be used for gases in the higher ppm ranges For measuring gases in the lower ppm ranges only stainless should be used If readings...

Page 14: ...You should now see flowing data on your monitor displaying the same information that is shown on the PPM1 or PPM2 Digital Display with a periodic update rate If you do not check your serial port setti...

Page 15: ...oard Select Value 1 to change ppmv NOTE PRESSING THE SPACE BAR SCROLLS THROUGH ALL THE AVAILABLE SELECTIONS PRESS ENTER TO SELECT THE ONE YOU WANT TO KEEP Press the Space Bar several times to see what...

Page 16: ...which corresponds to Dew Point in C Again use the Space Bar to scroll through the available options and then use Enter to select the one you want If you wish you can also change the measurement range...

Page 17: ...for each relay To enter the menu press C for C hange You will then see the window shown below In this example we would like to change Alarm 2 Note that it is now set for Dew Point in C and the relay s...

Page 18: ...h to We will leave the default F range of 50 0 to 0 0 Next we will program the alarm relay to open when the measurement is above 25 degrees and close when the measurement is below 24 degrees NOTE IT I...

Page 19: ...19 Next we are asked if we want to Save the Changes and we press Y for Y es Now we can select X for eX it and then A to A ccept the changes...

Page 20: ...outputs In order to utilize this capability connect the meter in series with one of the analog outputs and the load Observe the correct meter polarity Press the Escape key on your keyboard which will...

Page 21: ...for E xit and the Hyperterminal window will return to the default mode showing continuous updates 5 4 6 MODIFYING THE RS 232 PORT PARAMETERS Two RS 232 Port parameters may be changed in the field You...

Page 22: ...seconds which is a convenient rate to watch data updating on the computer monitor See the window on the right Once that is done you only have to press X for eX it and you will be back in the default...

Page 23: ...ferably should be made of sintered stainless steel A maintenance schedule should be set up to clean or replace the element periodically A clogged filter will retain moisture and slow down the system 6...

Page 24: ...one minute typical Hysteresis None Temperature Ambient 0 to 50 C Sample Gas 0 to 50 C Electrical Outputs Analog Outputs 2 4 to 20 mA Field Programmable Load 0 to 400 Digital Interface RS 232C bi direc...

Page 25: ...115 230Vac to 24Vdc Nominal Operating Time 8 to 10 hours Standard Cable 6 feet 1 8 meter Dimensions Carry Case LWD 19 in 48cm X 14 in 36cm X 7 in 18cm Weight Entire System 15 lbs 6 8kg Available PPM2...

Page 26: ...26 8 0 APPENDIX 8 1 PPM1 MOUNTING DIMENSIONS...

Page 27: ...27 8 2 SENSOR DIMENSIONS...

Page 28: ...nual specifications or other literature ii the items have not been misused or abused in any manner or repairs attempted thereon iii written notice of the failure within the warranty period is forwarde...

Page 29: ...29 Notes...

Reviews: