76
Introductions
A correction factor can be applied to the data collected with the EDC unit to account for variances in
gravimetric readings.
Calculating a Correction Factor
The correction factor is calculated by dividing the Gravimetric reading by the EDC unit reading.
Applying a Correction Factor
Follow the steps in the table below to apply a correction factor to all data points in the current project
folder.
1.
Select the 2
nd
Scale=
with a box where you can type in your scale factor.
2.
Type in the Scale factor.
3.
After the scale factor is entered press enter. All data points in the project folder have been
multiplied by the correction factor.
Removing the Correction Factor
Follow the steps in the table below to remove the correction factor from the data points in the project
folder.
1.
Select the 1
st
Scale=
under the Dataset Scale Factor
.
Data points should return to original state.